Maintenance
87
Maintenance
Contact for maintenance
Maintenance item list
To keep this product in good condition and operating at top performance, perform
maintenance regularly.
Table 29 Maintenance items
Before maintenance, drain the liquid from the extraction tank (refer to "Drainage mode
operation" (page 12)).
Manufacturer:
HORIBA Advanced Techno Co., Ltd.
31, Miyanonishi-cho, Kisshoin Minami-ku, Kyoto 601-8306, Japan
Item
Maintenance interval/frequency
Page
Rinsing the flow paths
When measurement is completed for the day
page 88
Inspecting the absorbent
sheet
Once a week
page 89
Cleaning the fan filter
Once a week
page 90
Washing the fan filter
Once a month
page 91
Washing the extraction tank
Once a year or if it becomes difficult to see the oil-water boundary
surface through the extraction tank window
page 92
Removing liquid from the air
hole of the extractor
If liquid gets into the extractor air hole
page 97
Replacing the water filter
If it takes longer than 60 seconds from the start of measurement until
settlement of a measured value
page 99
Drying the measurement cell If the measurement cell gets wet
page 102