 Preparation for operation

30

Preparation for operation

Operation check

Preparation

Perform the following check or work before turning ON the power.

Releasing optical unit lock screw

The optical unit in the detector is fixed by optical unit lock screws, to avoid possible vibration
or shock during transportation or mounting work.
When the detector mounting and wiring work are completed, remove the optical unit lock
screws (3 pcs), and replace them with the truss head screws with O-rings (3 pcs) supplied as
accessory. (Refer to Fig. 20.) 

Note

If the instrument is in operation with the lock screws fixed after the detector mounting, a correct
detection may not be possible. Further continuous use in this condition may damage the inside of
detector.
Before removing the detector, first confirm that the power is turned OFF, and replace the truss head
screws (three locations) with the lock screws retained. If transported with the truss head screws, it
may damage the detector due to shock during transportation.

Fig. 20  Release of optical unit lock screw

1.

Remove three optical unit lock screws.

2.

Replace the truss head screws with O-rings (M5).

Preparation

Attached round tray (for test)

Sample oil (olive oil, diesel fuel, kerosene, and others)

Tap water

Truss head screws with O-rings (M5)

Optical unit lock screws

Summary of Contents for LO-300

Page 1: ...Laser scanning type surface oil detector LO 300 Instruction Manual CODE GZ0000312982B...

Page 2: ...of the following cases none of the warranties set forth herein shall be extended Any malfunction or damage attributable to improper operation Any malfunction attributable to repair or modification by...

Page 3: ...EEE under the Directive 2012 19 EU and or waste batteries and accumulators under the Directive 2006 66 EC in the European Union The symbol might be put with one of the chemical symbols below In this c...

Page 4: ...ts are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency e...

Page 5: ...esult in death or serious injury This is to be limited to the most extreme situations This indicates a potentially hazardous situation which if not avoided could result in death or serious injury This...

Page 6: ...mmable gas explosive gas or inflammable dusts to avoid death or fire accidents CAUTION The weight of the detector is approx 11 kg main unit approx 9 8 kg accessory cable approx 1 4 kg Do not give stro...

Page 7: ...is a caution label to remind you that water invasion would destroy the instrument Warning Label This label mean Laser devices Label Description This messages are described in the following specificati...

Page 8: ...onnector name label This is a label to show you the maximum rating connector name and compliance to the CE marking Earth label This is a label to remind you the protective earth wiring is necessary fo...

Page 9: ...the upper lid of the indication box is removed due to maintenance work replacement of desiccants or fuse return it to the original state after the maintenance work and tighten the fixing screws 6 pcs...

Page 10: ...Disposal of the product When disposing of the product follow the related laws and or regulations of your country...

Page 11: ...s manual Note This interprets the necessary points for correct operation and notifies the important points for handling the product Reference This indicates the part where to refer for information Tip...

Page 12: ......

Page 13: ...12 Display box 13 Monitor screen 14 Status description 16 Operation 20 Screen configuration 22 Installation 24 Installation environment 24 Installing detector 25 Installation example using stand 25 Ch...

Page 14: ...check 55 Troubleshooting 56 Countermeasure at system error 56 Message list 58 Data 60 Specification 60 Parts list 62 Accessory 62 Option 62 Dimensional outline drawing 63 Detector 63 Detector stand o...

Page 15: ...water surface Signal processing by the built in CPU realizes the area detection function as well as removal of the noise due to ambient light Area detection function Alarms caused by reasons other tha...

Page 16: ...ter the small size of trunk covers large detection area from a distance Moreover this system can measure the oil covering ratio by the reflection time scanning The laser scanner module rotating around...

Page 17: ...tting and when the continuous judgment time passes If the detection area is set to 10 or more detection according to the size of the detection area is possible Detection level error When setting the d...

Page 18: ...om 0 to 90 Ensure the following Detection area WARNING Detection area ALARM Set the detection area by referring to the value of detection level monitor in the test mode Fig 2 shows an example of area...

Page 19: ...f 20 is continuously suspended on the detection area the threshold of alarm is set to 50 and the threshold of warning is set to 30 Monitoring status after setting change Check that the oil film alarm...

Page 20: ...by a power consumable apparatus nearby Noise on the power line 100 V AC for the detector Table 2 Contact output Detector status LCD Display Standard Oil film alarm AL contact System error ER contact...

Page 21: ...tion less With reflection less object such as debris LOW LEVEL 4 mA to 5 mA 4 mA to 5 mA 3 mA Continues monitoring With high reflective material as foiled aluminum HIGH LEVEL 20 mA or more 20 mA or mo...

Page 22: ...the threshold of the detection area Continuous judgment time setting CONT TIME Sets the required continuous judgment time for the alarm output Contact output setting RELAY OUT Selection of oil film wa...

Page 23: ...is invalid Continuous judgment time setting Warning 10 s Continuous judgment time setting Alarm 10 s Contact output setting Alarm The output of ALARM or WARNING can be selected Contact output hold ti...

Page 24: ...limit error Notification of laser replacement Lower limit error Normal return from oil film alarm oil film warning upper limit error lower limit error Temperature error Reference To confirm the histo...

Page 25: ...en directly Tip During the following status the back light does not turn OFF automatically Also when the following status occurs while turning OFF the back light the back light automatically lights up...

Page 26: ...name of the main body Top Front The rear of display box I O connector Side Power input Signal contact output Front lid of display box Handle Monitor 3 5 inch LCD with touch panel Note Remove the fron...

Page 27: ...nt lid of display box If the finger is held up during the OFF state the back light lights again CPU error LED CPU_ERR It is a red LED pilot light The light turns ON when the CPU has any operation erro...

Page 28: ...most grid is colored Detection area Displays the rate of oil film in the unit of 10 When the detection area is 20 two grids from the left are colored Detection level threshold Displays the threshold a...

Page 29: ...ing or on the Test mode screen for ten minutes or more Error message at startup Unexpected failure may occur when the following message is displayed at startup In this case contact us Factory preset d...

Page 30: ...uous judgment time passes For setting the judgment criteria refer to Detection level setting page 35 Detection area setting page 36 and Continuous judgment time setting page 37 The screen when the oil...

Page 31: ...Oil film alarm screen Upper limit error HIGH LEVEL The upper limit error occurs when the detection level exceeds the upper limit setting for 30 seconds fixed or more For setting the judgment criteria...

Page 32: ...The screen when the lower limit error occurs is as follows Fig 11 Lower limit error screen Instrument error SYSTEM ERROR The instrument error occurs when error such as Motor error or Temperature error...

Page 33: ...s displayed Also when the laser replacement time approx 1460 days has elapsed a system error contact is output Monitoring will continue but please replace the laser as soon as this message is displaye...

Page 34: ...hanges by touching the button in the above case WARNING or ALARM Increases or decreases the numerical value Depressing the button increases or decreases the value at a rapid rate ENT Enables various o...

Page 35: ...that Status change from oil film warning to oil film alarm the oil film alarm AL contact is output Status change from oil film alarm to upper limit error the instrument error ER contact is output Sta...

Page 36: ...UT Lower limit judge LOW LEVEL ERROR SETUP Refer to page 38 LEVEL ERR Multiplex transmission function MULTPLEX TRANSMISSION SETUP Refer to page 39 MULTI TRANS Set menu 2 2 MENU 2 2 Refer to page 35 NE...

Page 37: ...Refer to page 51 ANALOG OUT Contact force ful output RELAY OUTPUT Refer to page 52 RELAY OUT Water level adjustment RELAY SIGNAL OUTPUT SETUP Refer to page 53 REF INT CALIB Detection area monitoring...

Page 38: ...ody Place free from influence of inflammable corrosive or explosive gas Place with a wide water path width and where oil film does not easily gather right under the main body Use an optional shading h...

Page 39: ...ter surface 1 max 6 m The detector is ideally to be installed from the distance of approx 1 5 m For lightning surge prevention ensure to ground the cabinet and the stand For preventive measures of rai...

Page 40: ...with a handle Turn around the display box It is possible to change it by the pitch by 30 degrees Close the upper lid slowly and fix the screws M4 6 pcs securely with a torque wrench Tightening torque...

Page 41: ...t 4 mA to 20 mA Contact output Error ER Contact output Oil film alarm AL Unused Power cable 5 m accessory Ext Monitor connection Signal contact output cable 5 m accessory Crimped terminal indication W...

Page 42: ...he shield line of the accessory cables and the extension cable in the relay box Arrange the shield line SG of the signal and contact output cables on the receiving side to OPEN To extend the signal co...

Page 43: ...ng surge measures SG S S LO 300 Oil Laser Scanning type Oil Film 4 mA to 20 mA GND Oil film alarm AL contact NC NO OR CO CO 100 V AC U V GND G System error ER contact NC NO OR CO CO Housing GND GND Ho...

Page 44: ...trument is in operation with the lock screws fixed after the detector mounting a correct detection may not be possible Further continuous use in this condition may damage the inside of detector Before...

Page 45: ...xample when the detector normally monitors the water surface Detective operation test The operation test is executed by three conditions A No water B With water C With oil as in Fig 22 Fig 22 Detectiv...

Page 46: ...e screen At this time check the contact of instrument error refer to Table 6 Tip When the lower limit error is invalid NORMAL RUN is displayed on the screen Note that this is a normal operation B Test...

Page 47: ...the oil film Depending the oil type oil film may be hard to be generated on the water surface if this happens pour concentrate or water to the oil to perform the test Check the result of operation te...

Page 48: ...ing change that registration has not been completed will become invalid For the operations on screens refer to Touch panel operation page 20 and for the messages refer to Message list page 58 Set menu...

Page 49: ...COLOR Displays Color setting page 43 FACTORY Displays Factory default setting page 47 LM ALARM TIME Displays Laser replacement Contact output time setting page 48 BACK Displays Set menu 1 2 page 34 Na...

Page 50: ...e to set the detection area as follows WARNING detection area ALARM Name Description Detection area warning Setting range 0 to 90 Touch keys to change the setting The value can be set in each 10 Detec...

Page 51: ...white the background color of the digit of 100 is blue 4 Turn the value to 2 by keys 5 Touch the digit of 1 for WARNING The background color of the numerical value turns to white the background color...

Page 52: ...act output page 6 When the optional extended module is incorporated both WARNING and ALARM become the selected state Contact output hold time The setting range is 0 seconds to 99 seconds or infinity A...

Page 53: ...sion function touch ON or OFF on the screen When ON is selected the multiplex transmission function is valid When OFF is selected the multiplex transmission function is invalid Setting to ON allows cu...

Page 54: ...or judgment setting is OFF please set the lower limit level to 4 mA Normal detection level may not be output Name Description Upper level Any value exceeding this value will be judged as upper limit e...

Page 55: ...ue to 20 by keys 7 Touch Hour The background color of the value turns to white the background color of Day is blue 8 Turn the value to 17 by keys 9 Touch Minute The background color of the value turns...

Page 56: ...ntrast adjustment This screen enables to adjust the contrast of LCD display Fig 36 Contrast adjustment screen Name Description Contrast ratio key The contrast ratio rises key The contrast ratio lowers...

Page 57: ...itle Font Font color of the title Button Font Font color of the button Button Background Background color of the button Color Touch keys and select the color to set The colors that can be set are blac...

Page 58: ...re screen common to all screens Example Fig 38 Screen Background LEVEL AREA Bar Indicates the color of detection level bar and detection area bar on the monitoring screen in Fig 39 Fig 39 LEVEL AREA B...

Page 59: ...Indicates the color of parameters 1 and 2 on the monitoring screen in Fig 41 Fig 41 OIL WARNING Background Title Font Indicates the font color of the title of setting screen Example Fig 42 Title Font...

Page 60: ...es the font color of the button common to all screens Example Fig 43 Button Font Button Background Indicates the background color of buttons common to all screens Example Fig 44 Button Background Butt...

Page 61: ...ng touch Yes When the process is completed the screen returns to Set menu 2 2 page 35 To close the screen without returning to the factory default setting touch NO Tip After returning to the factory d...

Page 62: ...time when the laser replacement time approx 1460 days has elapsed Fig 47 Laser replacement Contact output time setting screen Name Description Hour Setting range 0 clocks to 23 clocks touch keys to ch...

Page 63: ...on alarm WARM Oil film detection warning HIGH Upper limit error LOW Lower limit error NORM Normal return from oil film alarm oil film warning upper limit error lower limit error TEMP Temperature error...

Page 64: ...and for the message refer to Message list page 58 Table 7 TEST mode menu TEST mode menu Touching TEST on the monitor display screen will display the test mode menu as follows Fig 49 TEST mode menu sc...

Page 65: ...he step below for forcible output of current 1 Set the output level by touching keys 2 Touch OUT the value set at the step 1 is used for current output Name Description Detection level Displays the de...

Page 66: ...side of screen Note The system error contact output for example NC CO is selected outputs approx 5 seconds delayed after an error is detected to avoid erroneous action ALARM Touching this button enab...

Page 67: ...p water at 1 m below the detector recommended 2 Touch keys to set the water level to 11 0 mA 3 Touching ENT will display the message Calibrating now on the screen enabling to start the adjustment 4 Wh...

Page 68: ...ion 1 To clear the laser operating days touch YES When the clearing operation completes the screen returns to Test mode menu To close this screen without clearing touch NO Note When attempting a reset...

Page 69: ...window may cause scattering or absorption of laser light disabling oil film detection Wipe off dirt using a glass cleaning agent Note If the dirt is hard to be removed use a neutral detergent Do not...

Page 70: ...receive laser light When the water level is detected the LOW LEVEL indication automatically disappears returning to a normal state When the LOW LEVEL indication is not required set the low limit erro...

Page 71: ...quency Place a white paper just under the optical window and check the rotation status of laser scanning light When the laser scanning light is not rotating contact us Instrument error Laser emission...

Page 72: ...n failed Displayed when the data registration is failed for data setting When the error persists even after the reset operation contact us 8 ERR Screen Title color Background color of the entire scree...

Page 73: ...e LOWER OIL LEVEL The condition lower limit level detection level is not satisfied Correctly set so that lower limit level detection level is satisfied 40 19 ERROR Calibration was failed Adjustment of...

Page 74: ...ating debris produces no reflection or abnormally strong reflection Contact output Number of outputs 2 Output type No voltage contact output Contact type Relay contact SPDT 1c Output capacity 250 V AC...

Page 75: ...re range 10 C to 50 C without freeze Operating humidity range Relative humidity 5 to 95 without condensation Compatible standards CE marking EMC Directive EN61326 1 Low voltage Directive EN61010 1 EN6...

Page 76: ...r release of optical unit lock screw Fuse 2 A 2 Knob M 5 25 3 Optical unit lock screw Instruction manual LO 300 1 Name Remark Detector stand SS Stand only made of SS with mounting bracket Detector sta...

Page 77: ...ne drawing Detector Detector stand option 90 25 120 25 185 30 99 448 547 20 60 255 100 Mounting bracket 4 M8 Unit mm 380 460 400 460 520 230 400 520 3 12 100 60 300 1000 90 60 4 M 8 Drain 20 90 90 90...

Page 78: ...Data 64 Mounting bracket option Shading hood 550 mm option 100 60 C 1 0 125 90 50 40 60 90 30 C 1 0 15 4 9 4 9 113 Unit mm 672 5 550 220 260 Approx 122 5 Unit mm...

Page 79: ...ion Steam protection hood option 572 5 450 220 260 Unit mm Approx 122 5 Steam protection hood attachment type Unit mm 500 185 260 Purge air inlet union Applicable tube size Outer diameter 6 Applicable...

Page 80: ......

Page 81: ...http www horiba adt jp Headquarters 31 Miyanonishi cho Kisshoin Minami ku Kyoto 601 8306 Japan TEL 81 75 321 7184 FAX 81 75 321 7291...

Reviews: