Other settings
74
■
Other settings
●
Printing measured values and calibration data
Print out the measured value or calibrated value displayed on the instrument, or the
measurement data or the calibration data saved in the instrument. If the repeatability is
inspected, the inspection data is printed out with the calibration data.
Pressing the
key starts printing during displaying the data you want to print. Use the
printer cable to connect the printer unit with the instrument beforehand. If the automatic
data memory is set to "ON", you cannot print out in the measurement mode.
Reference
For details of how to display measurement data and calibration data, refer to the respective
section.
"Displaying saved data" (page 50)
"Displaying the latest calibration and inspection data" (page 56)
Refer to the following table for the pH (two point calibration) and dissolved oxygen.
< The data saved in internal memory >
Printout format
Description
Memory Num
:0001
Data number
Date
: 2016/01/01
Measurement date
Time
: 09:00
Measurement time
Channel
: 1
Measurement channel
pH
: 5.22
Measured value and
out of the measurement range or displaying
range mark
Over: OR
Under: UR
Temperature
: 25.0°C MTC
Temperature value, temperature setting
and out of the measurement range or
displaying range mark
Over: OR
Under: UR
Inst. model
: LAQUAact-PD110
Instrument model
Inst. SN
: KL1TSE10
Instrument serial number
Elect. status
: GOOD
Electrode status based on calibration result
Memory Num
: 0001
Data number
Date
: 2016/01/01
Measurement date
Time
: 09:00
Measurement time
Channel
: 2
Measurement channel
DO
: 1.03 mg/L
Measured value
Temperature
: 25.0°C MTC
Temperature value and temperature setting
Inst. model
: LAQUAact-PD110
Instrument model
Inst. SN
: KL1TSN36
Instrument serial number