General settings
58
■
General settings
●
Setting the auto stability and auto hold function
This instrument has auto stability mode and auto hold mode.
• Auto stability mode (displayed as AS)
When the criterion for stability judgment is fulfilled during measurement, the component
icon lights and the measured value is fixed. Once the value deviates from the stability
judgment criterion, the measured value is released and displays instantaneous value.
• Auto hold mode (displayed as AH)
When the criterion for stability judgment is fulfilled during measurement, the component
icon and the HOLD icon light and the measured value is fixed. To release the measured
value, press the
key. Once a measured value is fixed, the measured value is not
released automatically even when the state deviates from the stability judgment criterion.
• Setting procedure
1. Press the
key to enter the setting mode.
2. Press the
keys to select "GEN" (general setting) and then press the
key.
3. Press
the
keys to select "MEAS" (measurement setting) and then press
the
key.
key
key
key
key