Measurement setting
53
●
Changing the calibration method
Select automatic calibration or manual calibration for the calibration method.
1. Press the
key to enter the setting mode.
2. Press the
keys to select "COND" (conductivity setting) and then press the
key.
3. Press the
keys to select "AUTO CAL" (automatic calibration setting) and
then press the
key.
4. Select "ON" to set the automatic calibration, or select "OFF" to set the manual
calibration. And then press the
key to confirm.
To return to the measurement mode, press the
key.
Note
When the setting of calibration method (auto or manual) is changed, the cell constant is
initialized. Make sure to enter the cell constant or to perform calibration again.