![horiba LAQUA-PH1500 Instruction Manual Download Page 36](http://html1.mh-extra.com/html/horiba/laqua-ph1500/laqua-ph1500_instruction-manual_2162253036.webp)
pH Setup
19
■
Setup
This section describes all the setup functions available in LAQUA PH1500 series handheld
meters.
● P1 pH Setup
Using P1 pH setup function of the meter, you can:
• Select buffer standard
• Set calibration alarm
• Erase calibration data
To set the pH functions, follow the procedure detailed below:
Prerequisites
Switch on the pH meter.
• Default buffer setup is
BUFF USA
. You can change it to
BUFF NIST
or
BUFF DIN
if
required.
• Calibration alarm setup option must be used to avoid “Calibration interval alarm error” on
page 37 . You can set the calibration alarm for ---- day to upto 90 days, where ----
indicates “no calibration alarm” has been set.
• Erasing previous calibration data is recommended for accurate calibration. Default setup is
NO
but to erase the calibration data, you need to change the setup to
YES
.
To return to the measurement mode, press the
key.
Note
Tip
Summary of Contents for LAQUA-PH1500
Page 1: ...Instruction Manual Benchtop pH ORP Meter LAQUA PH1500...
Page 2: ......
Page 4: ...MEMO II...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 23: ...Product Overview 6...
Page 49: ...General Setup 32...
Page 53: ...Maintenance and Storage 36...
Page 63: ...Meter Specifications 46...