Conductivity Calibration
15
● Conductivity Calibration
Calibration is necessary for accurate electrical conductivity measurement. To perform
conductivity calibration, follow the procedure detailed below:
Prerequisites
• Clean the conductivity electrode with DI (deionized) water and wipe it with tissue paper.
• Switch on the meter and plug in the conductivity electrode.
• Prepare standard solution required for calibration.
• Press the
key to keep the meter in
COND
mode.
• Dip the conductivity electrode in the standard solution until the hole at the upper part of the
electrode is immersed.
• Salinity, TDS, and resistivity of a sample solution are calculated from the measured value
of conductivity.
• In conductivity calibration mode, the default calibration method is auto calibration. If you
like to change it to manual calibration method, refer “P1.3 Calibration Mode Setup” on
• For second or multiple point calibration, clean the conductivity electrode with DI water and
follow the same procedure.
• If you are performing multiple point calibration, calibrate to the lowest conductivity first and
then move to increasing conductivity values. This minimizes cross contamination.
• To abort an ongoing calibration process at any point of time, press the
key.
Note
Tip
Summary of Contents for LAQUA-PC1500
Page 1: ...Instruction Manual Benchtop pH ORP Conductivity Meter LAQUA PC1500...
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 24: ...Product Overview 6...
Page 46: ...pH Setup 28...
Page 54: ...Conductivity Setup 36...
Page 58: ...TDS Setup 40...
Page 65: ...Data Setup 47 Conductivity Mode key key key key key key...
Page 67: ...Data Setup 49 Conductivity Mode key key key key key key key...
Page 69: ...Data Setup 51 Conductivity Mode key key key key key key key...
Page 72: ...General Setup 54 Conductivity Mode key key key key key key key...
Page 74: ...General Setup 56 Conductivity Mode key key key key key key key...
Page 76: ...General Setup 58 Conductivity Mode key key key key key key key...
Page 78: ...General Setup 60 Conductivity Mode key key key key key key key...
Page 90: ...Printout Summary 72 Appendix 2 pH mV Relative mV Printer Format Measurement...
Page 91: ...Printout Summary 73 Resistivity Conductivity...
Page 92: ...Printout Summary 74 Salinity TDS...
Page 93: ...Printout Summary 75 Printer Format Data Log...
Page 94: ...Printout Summary 76 Conductivity pH Salinity Printer Format Calibration...
Page 99: ......