Basic operation
11
●
Entering numeric values
When entering numeric values to make various settings and set a calibration value, you
can change the selected digit using the
keys and increment or decrement the value
(0 to 9) using the
keys.
●
Saving measurement data in the internal memory
Up to 999 data items measured by the instrument can be stored in the internal memory.
Saving the measurement data is possible only when the instrument is in the
measurement mode.
1. Press the
key while the data to save is displayed.
The saved data is displayed for two seconds, and then the display returns to the
previous screen automatically.
Note
If the data saved reaches 999, an error occurs and "ERROR No. 0010" is displayed. Copy
or transfer necessary data to a PC and delete the data from the memory (refer to “ Deleting
all saved data ” (page 55)).
key
key
key