● Maintenance and Storage of Electrode
This section describes the procedures for maintenance and storage of conductivity electrode.
● Cleaning the Conductivity Electrode
Always clean the electrode with deionized water after every measurement. When the response
is slow or residue from the sample adheres to the electrode, use the appropriate method below
to clean the electrode, and then clean again with deionized water.
● Conductivity Electrode Storage
If the electrode is stored in a dry state, the cell constant will change. Store with the black
electrode part immersed in deionized water, or with the protective cap filled with deionized
water and attached to the electrode.
To store the electrode for a long period, clean it well and attach the protective cap filled with
deionized water.
Type of dirt
Cleaning solution
General
Diluted neutral cleaning solution
Inorganic substance
Ethanol (keep the ethanol away from plastic parts)
Scale that formed during
long term storage
A commercially available scale remover (neutral cleansing
solution for kitchen use, etc.) diluted by a factor of 100. If this
does not remove the scale, use diluted solution that contains
oxygen bleach (sodium percarbonate) or chlorine bleach
(sodium hypochlorite).
Summary of Contents for LAQUA-EC1500
Page 1: ...Instruction Manual Benchtop Conductivity Meter LAQUA EC1500...
Page 2: ......
Page 4: ...II MEMO...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 24: ...Product Overview 6...
Page 46: ...Conductivity Setup 28...
Page 50: ...TDS Setup 32...
Page 64: ...General Setup 46...
Page 76: ...Printout Summary 58 Appendix 2 Printer Format Measurement Conductivity Resistivity Salinity...
Page 77: ...Printout Summary 59 TDS Printer Format Data Log...
Page 78: ...Printout Summary 60 Printer Format Calibration Conductivity...
Page 79: ......