Chapter 3 COND (conductivity) Measurement
34
HORIBA
Chapter 3 COND (conductivity) Measurement
3.1 COND Calibration
This section describes the procedures to set the conditions of COND calibration.
Set the condition of temperature compensation before COND calibration according to "3.2
COND Measurement Setting" (P.38).
The cell constants of COND electrodes are different.
When using the conductivity electrode for the first time, set the cell constant written on the
electrode into the instrument before use according to "3.2.1 Cell Constant Setting" (P.38).
3.1.1 Automatic Calibration Setting
1.
Press the CAL key.
2.
Tap
and tap "CH2 CAL SET".
A screen to select Cell Constant and Auto
Calibration is displayed.
3.
Tap
on the right of the Auto Calibration item.
4.
Tap
.