Data
70
Data
Specifications
Converter
Product name
Turbidity meter
Model
HU-200TB-W Industrial-use turbidity converter
Turbidity sensor type SS-120-W Industrial-use turbidity detector
Measurement range
Kaolin
0 to 1000 (degrees) (Display range: 0 to 1100 (degrees)) or 0 to 1000 (mg/L)
Formazin (NTU) 0 to 2000 (NTU) (Display range: 0 to 2200 (NTU))
PSL
0 to 100 (degrees) (Display range: 0 to 110 (degrees))
Transmission output
setting range
Kaolin
Arbitrary setting within the range from 0.00 to 10.00 (degrees), 0.0 to 100.0
(degrees), 0 to 2200 (degrees)
Formazin (NTU)
PSL
Display resolution
0.01 degrees (Degree range: 0 to 10 (degrees))
0.1 degrees (Degree range: 0 to 100 (degrees))
1 degrees (Degree range: 100 to 2200 (degrees))
Selectable between fixed range (decimal point) and automatic range changeover
Performance
Repeatability
Within
2
of the reading or
0.2 degrees, whichever is larger (checked using
the solid span bottle)
Linearity
The deviation at the mid-range point of the span calibration value is within
2
of the span calibration value or
0.04 degrees, whichever is larger
Accuracy
Within
2
of reading
Transmission output
Number of current outputs: 2
Output type
4 mA to 20 mA DC, input/output isolation type
Load resistance 900
(max.)
Linearity
Within
0.08 mA (limited to the current linearity)
Repeatability
Within
0.02 mA (limited to the current repeatability)
Error output
Provided with burnout function (3.8 mA or 21 mA)
Hold function
Selectable between the latest value hold and the preset value hold
Contact output
Number of output points: 6
Output type
No-voltage contact output
R1
, R2
Contact type
Relay contact, SPST (1a)
Contact capacity
250 V AC 3 A, 30 V DC 3 A (resistance load)
Contact function
Selectable from among the upper limit alarm, lower limit
alarm, during transmission output hold, cleaning output and
automatic calibration (closed at alarm operation, open under
normal conditions and in a power-off state)
Description of alarm
Setting range: Within turbidity measurement range
Delay time: 0 to 600 (s)
R3
Contact type
Relay contact, SPST (1a)
Contact capacity
30 V DC 1 A (resistance load)