![horiba HU-200TB-W SS-120-W Instruction Manual Download Page 33](http://html1.mh-extra.com/html/horiba/hu-200tb-w-ss-120-w/hu-200tb-w-ss-120-w_instruction-manual_2162226033.webp)
Installation
21
Failure output
If you want a break signal for power OFF or abnormality, use the contact with CF and NCF.
When this equipment is normal (not malfunctioning), the CF-NOF contact is opened, and the
CF-NCF contact is short-circuited.
When the power supply is disconnected, the CF-NOF contact is short-circuited.
WARNING
When normal
Abnormal condition
(FAIL occurrence)
Power-OFF
CF-NOF
Opened
Closed
Closed
CF-NCF
Closed
Opened
Opened
NOF
FAIL
CF
NCF