Installation
20
Contact output R1 to R3
The terminal screw for the contact output are M4 type.
The applicable electric wire is of 0.75 mm
2
to 5.5 mm
2
(AWG18 to AWG10).
For wire connection, remove the black resin terminal cover for the power terminal by
pulling it forward and raising it, or by detaching the hinge.
Contact capacity
Tip
Set an assignment of the contact in the hold mode.
For each contact, a function can be assigned from among the upper/lower limit, during cleaning,
during calibration, hold state, at the completion of cleaning, and at the completion of calibration.
Resistance load
R1, R2 Within 250 V AC 3 A or 30 V DC 3 A
R3
Within 30 V DC 3 A
R3
R1
R2