Installation
11
Installation
Installation environment
To keep the HU-200TB-W stable and reliable for use, install it in a place where the following
conditions are satisfied.
Converter
Well-ventilated location.
Ambient temperature in the range of
20
C to 55
C.
Not exposed to direct sunlight.
Not exposed to direct radiant heat.
Ambient relative humidity in the range of 90
without dew condensation.
The converter is not splashed with water or chemicals.
Mechanical vibration is rare.
There is sufficient space for maintenance and wiring.
Dust and corrosive gases are not present.
The influence of electromagnetic fields is rare.
The altitude is 2000 m at maximum.
The range of power supply voltage fluctuation is within 10
of the rated voltage.
Detector
Sensors can be checked and maintained.
Spilled water does not adversely affect the location.
Utilities are available for zero water.
Measured liquids should be those that will not damage the portion of the detector that
comes into contact with the liquid.
Samples do not freeze.
For an automatic cleaner, a 5 m cable can be connected between the automatic cleaner
and the converter.
Note
If measurements are made without depressurizing a pressurized sample, water may leak from the
pressure releasing hole. This is not an abnormality.