![horiba GA-370 Instruction Manual Download Page 78](http://html1.mh-extra.com/html/horiba/ga-370/ga-370_instruction-manual_2162216078.webp)
6 Troubleshooting
62
6.2.12 Settings cannot be changed
6.3
Remedial actions when events occur
Events are classified into three types: information, alarms, and cautions.
If an event is occurring, refer to "3.6 Event check" (page 34) and then check the details of the
event and take the corresponding remedial action as described below.
In the case of symptoms for which a probable cause is not described or a problem that is not
resolved even after taking the described remedial action, contact HORIBA.
6.3.1 Types of information and remedial actions
Table 28 Types of information
*This is not displayed in the EVENTS screen. Check the EVENT HISTORY screen.
Cause
Remedial action
Reference
Key lock is set.
Unlock the keys.
Others Contact
HORIBA
or local representative.
―
Name
Description
Check details
Remedial action
Contact output
SYSTEM
START*
The analyzer system
has started.
―
―
―
SHUTDOWN*
A system shutdown
was performed.
―
―
―
MEMORY
An error has occurred
with the HORIBA
service memory.
―
The HORIBA service
memory needs to be
replaced. Contact
HORIBA or local
representative.
There is no effect on
concentration value
measurement while
this is occurring.
―
DATETIME
CHANGE*
The time was
changed.
―
―
―
FILTER
The passed time of the
filters has exceeded
the recommended limit
time.
―
Replace the filters.
―
PACKING
The passed time of the
packing has exceeded
the recommended limit
time.
―
Replace the packing.
―