DO Electrode (9520-10D) Operation Manual 

� 

� 

Thank you for purchasing a HORIBA DO electrode. 

Read this Operation Manual carefully before using the electrode. 

©  Copyright HORIBA,  Ltd.  1997 

Cautions during handling 

Ref erence Only 

• The D-20 Series of pH meters can be used in combination with this electrode to provide water-resistant construction (conforming to IP-67). When

measuring, do no! immerse lhe cap, lead or connector in the sample.

• Do no! allow lhe electrode to come in contact with any hard surface.

• An extremely thin membrane is stretched across lhe DO tip and DO responsive par!. DO no! allow lhe membrane to come in contact with any hard

surface and do no! exert undue force on lhe membrane. 

• The internai solution of lhe DO tip is a strong alkaline. lf lhe internai solution comes in contact with your clothes or skin, flush immediately with water.

lf lhe internai solution comes in contact with your eyes, flush immediately with large amounts of water and seek treatment by a physician.

• When attaching lhe DO tip to lhe electrode body, check to make sure that lhe O-ring is attached. Attach lhe DO tip securely to lhe electrode body.

• The electrode connector requires a high degree of insulation. Do no! allow lhe connector to come in contact with water or dirty hands.

• DO tips that are no longer usable should be disposed of as industrial waste.

Contents 

• Electrode body

• DO tip (#7541)

• Sensor adapter (common taper fit for$ 19)

1 pc. 

1 pc. 

1 pc. 

1 pc. 

1 pc. 

• Stirrer

• Stirrer casing

• Operation Manual

1 copy 

Specifications and names of parts 

• Model: 9520-10D

Specifi­

cations 

Names 

of parts 

Measurement principies

Measurement range 

Usable temp. range 

Wetted materiais 

Dimensions: 

Electrode length 

Ext. diam. of wetted par! 

Lead length 

Stirr

¼

er 

cas1ng 

O-ring

DO tip 

(#7541) 

Stirrer 

Membrane galvanic cell 

O - 19.99 mg/1 

O - 45 ºC 

PVC, POM, epoxy, FEP, CR, silicon 

rubber 

190 mm (incl. cap) 

15 m 

1  m 

Electrode 

connector 

onector 

pin 

Cap 

Membrane 

Preparations 

4.1  Connecting to pH meter 

1  Inser! lhe electrode connector 

into the connector por! sleeve 

on lhe meter, after aligning with 

lhe pin. Do not inseri lhe con­

nector unless it is aligned prop·· 

erly with lhe connector port. 

C2 

Press lhe electrode connector 

into lhe connector por! on lhe 

meter, while turning the connec­

tor to lhe right. 

3 Slide lhe connector cover over 

lhe connector. Then, screw lhe 

cover to lhe right until is comes 

in light contact with lhe meter 

case. 

Inseri lhe temperature connec­

tor into the jack on lhe meter. In­

ser! lhe connector firmly, until 

lhe O-ring on lhe connector can 

no longer be seen. 

Note 

►►► 

With a D-25 pH meter, lhe C2 and T2 connectors are used 
for lhe DO electrode. The C1 and T1 connectors are re­
served for pH or ORP electrodes, only. 

4.2  Preparing electrode 

� 

Caution 

Chemical solution 

Highly concentrated potassium hydroxide (KOH) is 

used in lhe internai solution of lhe electrode. 

lf the internai solution in lhe electrode comes in con­

tact with your hands or skin, wash immediately with 

water. lf lhe internai soluti.on comes in contact with 

your eyes, flush immediately with large amounts of 

water and seek treatment by a physician. 

• The sensor adapter is used to make lhe container being used air-tight

during BOD measurement, when it is extremely important that lhe sam­

ple contained in a furan bottle during measurement not absorb any oxy­

gen from lhe air, or when storing a DO electrode in zero sample solution

inside a conical flask.

1  Attach lhe sensor adapter to lhe 

electrode body. 

-�

Electrode 

body 

Sensor 

adapter 

Remove the DO tip (#7541) 

from its packaging. 

Note 

►►► 

An O-ring is attached inside lhe sensor adapter, so push 
lhe sensor adapter in with a litlle force, while turning it. 

3 Remove the DO tip short sock­

et. 

Short 

socket 

Put lhe stirrer into lhe stirrer 

casing. 

� 

Stirrer 

�Uccec oasióg 

Push lhe DO tip securely into 

the electrode body, to attach it. 

DO tip 

Electrode 

body 

6 Put lhe stirrer casing onto lhe 

DO tip, so that it clicks into 

place. 

�� 

� Stirrer casing 

Note

►►► 

Do not dispose of the short socket that was removed. li is 
necessary when storing lhe electrode. 

Note

►►► 

When attaching lhe DO tip to lhe electrode body, make 
sure that lhe O-ring is attached to lhe DO tip. 

Points to remember when measuring 

(calibrating) 

• Air calibration

• Remove any drops of liquids from lhe membrane ai lhe tip of 

lhe DO electrode, by wiping away lhe liquid with soft tissue 

paper or similar material to make it dry.

• Avoid calibrating in places with extreme temperature fluctuation,

places exposed to the wind or rain, and places near heating

equipment.

• Do not grasp lhe holder during or immediately before/after cal­

ibration. (Body temperature adversely affects lhe stability of lhe

readings.)

Note

►►► 

Equalizing lhe temperatures of lhe DO electrode and lhe air 
is important in achieving proper air-calibration. lf possible, 

allow the DO electrode to sit in lhe open air for a short time 
(approx. 20 minutes) prior to calibration. 

• Standard solution calibration and sample

measurement

• Wash lhe electrode in pure (ion exchange) or tap water and dry

it with filter or tissue paper, each time prior to immersing it in lhe

solution.

• When immersing lhe electrode in lhe solution, be sure that lhe

temperature sensor is completely immersed.

• S!ir lhe DO electrode around in lhe solution, so that lhe temper­

ature of lhe electrode becomes lhe sarne as that of the solution.

• Make sure lhe membrane has no dirt or air bubbles on it.

• Measure lhe DO of the solution, making sure lhe flow speed on

lhe surface of lhe membrane is constant.

• lmmerse lhe electrode into lhe

solution at an angle, so air 

bubbles do not cling to lhe DO 

tip membrane.

• Start lhe stirrer, setting it for

approx. 1,000 rpm or more to 

stabilize lhe DO meter reading.

Temp. sensor 

Stirrer 

Maintenance 

• After finishing measurement, wash lhe DO electrode well in

pure (ion exchange) water, then keep it immersed in a conical

flask containing zero standard solution (approx. 5% sodium

sulfite solution). Leave lhe electrode connected to lhe meter. 

Note 

►►► 

lf lhe electrode is stored for a long period of time without 

being connected to lhe pH meter, it may be difficult to stabi­
lize lhe reading when preparing for to make measurements. 

Wash lhe DO electrode well 

lmmerse lhe electrode in a coni-

with pure (ion exchange) water. 

cal flask containing zero stan­

dard solution (approx. 5% so­

dium sulfite solution). 

Membrane 

Sensor 

adapter 

Zero standard 

solution 

• lf the electrode membrane is dirty, clean it by using soft cotton gauze ,

using care not to scratch lhe membrane. Then, rinse lhe membrane with

pure (ion exchange) water. 

• lf an errar occurs during calibration, replace lhe DO tip.

Note

►►► 

As a DO electrode is used, white sedimentation may devel­
op in lhe internai solution of the DO tip. This does no! affect 
performance. 

Storage 

• The electrode should, as a principie, be stored while remaining con­

nected to lhe meter. 

• To store lhe electrode without the pH meter, follow lhe below procedure.

1  Remove lhe electrode from lhe 

connector on lhe meter. 

Remove lhe DO tip from lhe 

electrode body. 

�-@; 

DO tip 

Electrode body 

Wash lhe electrode well with 

pure (ion exchange) water, then 

dry it with cotton gauze. 

Screw lhe short socket onto lhe 

DO tip, then store it in a coai, 

dark location. 

Note

►►► 

Avoid storing lhe electrode in a hot place or places with 
high humidity. Store lhe electrode indoors, out of direct 
sunlight. 

* "DO"  is  an abbrebiation of Dissolved Oxygen.

Technical questions regarding this product should be directed 

to lhe following HORIBA Customer Support Center. 