10 APPENDIX
92
10.3 Specification
Model
APNA-370
Measurement target
Nitrogen oxides (NOx, NO
2
, and NO) in atmospheric air
Measuring principle
Cross modulation type chemiluminescence method
Range
Standard
0 ppm to 0.1/0.2/0.5/1.0 ppm
Automatic range switching
Optional
Max. 5 ranges between 0 and 0.1/10 ppm, Maximum range ratio: 10
Minimum detection
sensitivity
For ranges of 0.2 ppm or less: 0.5 ppb (2
)
For ranges exceeding 0.2 ppm: 0.5% (2
) of the full scale
Reproducibility
(repeating accuracy)
±1.0% of the full scale
Linearity
(readout error)
±1.0% of the full scale
Zero drift
±1.0% of the full scale/day
±2.0% of the full scale/week (ambient temperature change: within 5°C)
Span drift
±1.0% of the full scale/day
±2.0% of the full scale/week (ambient temperature change: within 5°C)
Response rate
120 s or shorter (T
90
from the inlet)
Interference effect
Moisture 2.5%:
Zero ±2.0% of the full scale
Span ±3.0% of the full scale (for ranges of 0.2 ppm or less)
NH
3
1 ppm:
For ranges of 1 ppm or less: ±4 ppb
For ranges exceeding 1 ppm: ±1.0% of the full scale
Sample collection
rate
Approximately 0.8 L/min
Display
Measured value, alarm, time, alarm history, calibration history, etc.
Alarms
Zero calibration, Span calibration, Converter temperature, etc.
Input/output
0 V to 1 V
(2 lines of momentary and rolling average values or average values)
Contact input/output
(range, alarm, etc.)
RS-232C (optional)
Ambient temperature 5°C to 40°C
Ambient humidity
For lower than 31
C, the relative humidity must be 80% or lower.
For temperatures between 31
C and 40
C, the relative humidity must be below a linearly
declining range from 80% at 31
C to 50% at 40
C.
Altitude
3000 m above sea-level, or lower
Power source
100, 115 V ±10 V AC 50/60 Hz, or
220, 230, 240 V ±10 V AC 50 Hz (depending upon the specifications)
Power consumption Approximately 170 VA in steady state, approximately 220 VA in maximum
External dimensions 430 (W)
221 (H)
550 (D) mm
Mass
Approximately 21 kg
Connections
Sample inlet:
Joint for 6 mm O.D./4 mm I.D. Teflon tube
Calibration gas inlet:
Joint for 6 mm O.D./4 mm I.D. Teflon tube
Exhaust gas:
Joint for 6 mm O.D./4 mm I.D. Teflon tube