2 BASIC OPERATIONS
6
2.3
Basic Operation Flow
To perform operations, ensure that the installation, wiring, and piping connections have been
completed.
(Connect the external input/output as necessary.)
For the first use
*1: The default password is 1234.
Power ON
Turn ON the power.
“2.1 Start-up (Measurement Start)” (page 4)
↓
Setting
Unlock the keys*
1
Set the current time.
“6.5.1 Time adjustment” (page 54)
Set the start time, interval for calibration mode
or operation using the internal clock.
Set the calibration sequence (zero span time).
“4.3.3 Setting the AIC sequence” (page 23)
↓
Output setting
Set the analog output range (Fixed, Auto, or
External).
The default setting is "Auto."
Select a desired mode in accordance with your
use.
↓
Password change
The default value is 1234.
Change this value as necessary.
“6.6.3 Password setting” (page 61)
↓
Zero/span gas
connection
Connect the zero/span gas line to be used and
then check this connection.
↓
Span gas
concentration entry
Enter the concentration of the span gas to be
used.
“4.2.1 Entering the span gas concentration
value” (page 14)
↓
Calibration
Perform calibration automatically or manually.
“4.3 Automatic Calibration (AIC)” (page 16)
“4.4 Manual Calibration” (page 26)
↓
Measurement
Perform the continuous measurement.