ABX Micros
ES60
OT/CT
6–14
User Manual - RAB237AEN
Tab.6-2 Default values of XB alarms
6.5. Analysis running
◆
The "
Running
" screen allows to set analysis running modes.
In the "
Analys.
" tab:
1-
Press "
Running
" radio button.
2-
Press "
Edit
" button.
3-
Check the "
With RUO data
" check box to activate/deactivate RUO data. It is activated by default
(except for US version).
4-
Check the "
Automatic lock on analyser alarm
" check box to activate/deactivate instrument
locking on analyser alarms (reagent temperature, invalid startup). It is deactivated by default
(except for US version).
5-
Check the "
Automatic lock on reagent alarm
" check box to activate/deactivate instrument
locking on reagent alarms. It is deactivated by default (except for US version).
6-
Check the "
Automatic lock on QC alarm
" check box to activate/deactivate instrument automatic
locking on QC alarms (inter QC time limit or invalid QC). It is deactivated by default (except for US
version).
7-
Check the "
Display startup results
" check box to activate/deactivate displaying (at the end of
startup cycle) results of blank cycle runs during the startup. It is activated by default.
8-
Check the "
Print startup results
" check box to activate/deactivate automatic printing results of
blank cycle runs during the startup. It is activated by default.
9-
Press "
OK
" button.
6.6. Identification option
◆
The "
Identifi.
" screen allows to activate or not automatic identification. This option allows the
instrument to generate automatically sample ID.
In the "
Analys.
" tab:
1-
Press "
Identifi.
" radio button.
2-
Press "
Edit
" button.
3-
Check the "
Automatic identification
" check box to activate/deactivate automatic identification.
It is deactivated by default.
4-
Press "
OK
" button.
Alarm
Min. value
Max. value
MCV (µm3)
80
100
MCH (pg)
27
32
MCHC (g/dl)
32
36
Summary of Contents for ABX MICROS ES60
Page 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Page 2: ......
Page 6: ...User Manual RAB237AEN...
Page 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Page 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Page 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Page 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Page 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...