15
Electrode Preparation
Calibration
Measurement
After Measurement
R
insing the electrode
Make sure that no dirt and other foreign matters have
adhered to the electrode before measurement.
If dirt and other foreign matters have adhered to the
electrode, rinse the electrode as described in the following
procedure.
1
Rinse the electrode with pure water.
2
Wipe off the moisture.
Wipe off the moisture using soft and clean lint-free paper or cloth.
3
Make sure that there are no dirt and
crystals of internal solution adhered
to the area around the liquid
junction.
If dirt and crystals of internal solution are adhered, they
may affect the outflow of internal solution. Rinse the
electrode with pure water again.
View video
View video
Contamination
Note
Note