Electrode Preparation
Calibration
Measurement
After Measurement
C
alibration operation
Measure standard solutions in ascending order of ion concentration to create
calibration curves in the meter (
“Ammonia electrodes” (page 54)).
Perform calibration once every day or before measurement as appropriate.
Before operation
Be sure to fill with internal solution and rinse the electrode
before the calibration operation. (
electrode” (page 8), “Rinsing the electrode” (page 12))
Perform conditioning in the following cases:
Before the first use
Before use after storage for one week or longer
If the electrode is less responsive
After the replacement of the membrane cap
(
Note on page 21)
* :
Prepare 5 mol/L aqueous solution of sodium hydroxide, or use
the pH-adjusting ionic strength adjuster (referred to as pH-
adjusting ISA in the rest of this document) available as consum-
ables. (
Note
Note
Syringe for
adjusting pH
Standard
solutions
Items to be prepared
pH-adjusting ionic
strength adjuster
*