Dissolved oxygen measurement
33
• Setting salinity correction (default: 0.0 PPT)
Because dissolved oxygen is affected by the salinity concentration, you need to correct it
at measurement. Set the salinity concentration of the sample beforehand.
You can select arbitrary numeric value or sea concentration (35.0 PPT). If the measured
target does not contain saline matter, set to 0.0 PPT.
1. Press the
key to change the setting mode.
2.
Press the
keys to select the "SAL" (salinity correction setting) and then
press the
key.
3. Select the numeric setting display to set a salinity coefficient, or select "SEA" to
set a sea concentration. And then press the
key.
When the numeric setting display is selected, enter the salinity coefficient and press
the
key.
The setting range is 0.0 PPT to 40.0 PPT.
4. Press the
key to reflect the change.
key
key
key
Press the
key in the
numeric setting
display
Summary of Contents for LAQUAact D-75G
Page 2: ......
Page 22: ...M E M O...
Page 88: ...M E M O...
Page 94: ...M E M O...
Page 116: ......
Page 117: ......
Page 118: ......