background image

Dissolved Oxygen Calibration & Measurement

The 

 will start blinking until the reading stabilizes.

1. Rinse the DO electrode 

with clean water and blot 

dry with lint-free tissue. 

2. Hold the DO electrode still in clean air.

3. Press CAL button on the meter to switch 

to calibration mode.

4. Press ENT button to confirm the DO reading. To abort calibration, press 

MEAS button.

Perform air calibration in clean air at 

a location not subjected to dramatic 

temperature change, rain, or direct wind. 
Two calibration modes are available in the 

DO meter:

1. Saturated oxygen concentration mode (%)

2. Dissolved oxygen measurement mode 

(mg/L)

The default mode is the former. Press MODE 

button to switch to the latter.

Make sure that the membrane and 

temperature sensor of the DO electrode are 

fully immersed in the sample. 

5. Immerse the DO electrode in zero solution (5% sodium sulfite) and repeat step 

nos. 3 - 4 to perform zero calibration (optional).

6. Rinse the DO electrode with clean water and blot dry with lint-free tissue.

7. Immerse the DO electrode in sample.

Make sure that the membrane and temperature sensor of the DO electrode are fully immersed in the 

zero solution. 

In step no. 3, CAL button must be pressed twice so that the meter will switch to zero calibration mode. 

Quick Guide

LAQUA 200 Series 

DO210/220 & PD210/220 

Handheld Meters

www.horiba-laqua.com

The meter will switch to measurement 

mode and the electrode icon will 

appear on the screen.

DO210

Press

Press

There should be no water droplets on 

the membrane and temperature sensor.

Reviews: