![Horiba Scientific 6561S-10C Instruction Manual Download Page 32](http://html1.mh-extra.com/html/horiba-scientific/6561s-10c/6561s-10c_instruction-manual_2162341032.webp)
20
S
tandard solutions
The following standard solutions are used for calibration and measurement.
1000 mg/L F
-
standard solution
Use the standard solution included in the package or available as
consumables by HORIBA Advanced Techno.
100 mg/L F
-
standard solution
Use the standard solution included in the package or available as
consumables by HORIBA Advanced Techno.
10 mg/L F
-
standard solution
Prepare this by pouring 50 mL of the 100 mg/L F
-
standard solution into a
500 mL volumetric flask and filling up to 500 mL with pure water.
Summary of Contents for 6561S-10C
Page 1: ...Instruction Manual Combination type Fluoride Ion Selective Electrode 6561S 10C ELECTRODE...
Page 8: ......
Page 20: ...8...
Page 30: ...18...
Page 54: ...42...
Page 60: ...48...
Page 70: ...58...
Page 87: ...75 Material External dimensions Unit mm 16 5 O D 18 85 150...