UVAHAND LED
16
# 073350DE-EN b
5.
Servicing, Maintenance and Cleaning
5.1.
Service
The following servicing tasks must be carried out on the UVAHAND LED:
Regularly (depending on the operating conditions):
Check the safety glass plate for dust and other contamination.
Clean the safety glass plate as necessary and replace the safety glass
unit if it is damaged.
5.2.
Servicing
Replacing the safety glass unit
The safety glass unit consists of the safety glass plate and the safety glass
frame.
The safety glass unit must be replaced immediately if the safety glass plate is
damaged.
DANGER!
Under no circumstances may the unit be operated with a
damaged safety glass plate or without a safety glass plate.
Failure to follow these instructions can result in damage to
the skin and eyes caused by the UV radiation emitted!
DANGER!
Before opening the equipment, switch it off and disconnect
from the mains. There is danger of direct or indirect contact
with electricity!
WARNING!
The UVAHAND LED emits high-intensity radiation. There is a
danger of both fire and burns. Wear corresponding protective
equipment.
Allow the unit to cool down for 10 min.
WARNING!
Use only spare parts manufactured by the Dr. Hönle AG.
Safe operation of the UVAHAND LED cannot be guaranteed
if parts from other manufacturers are used.
NOTE!
To order replacement parts, see "Ordering data for units,
replacement parts and accessories; Replacement Parts."
Service