Replace an Analyzer
The Vertex Edge rack is designed for quick replacement of major components. You may replace
the Chemcassette analyzer while other analyzers continue to monitor.
WARNING: Do not connect or disconnect anything from the Power Distribution Unit (PDU)
while energized.
1. Turn off monitoring mode for the Analyzer to be replaced. In the left navigation panel, tap
Runtime Control
2. Tap on the Analyzer Monitor tab.
3. In Runtime Operation, tap on the Analyzer operation tab.
4. Tap Release and pull out the analyzer as far as possible
5. Tap the open the gate command if you wish to remove the Chemcassette. After removing the
Chemcassette put the optics protection cards in each optic block.
6. Close the gate.