4. Start up
4.6. Calibrating the transmitter
Revision 2
SmartLine Wireles User's Manual
47
Procedure for field calibration of Honeywell compound characterized SmartLine Wireless
Transmitters
This procedure applies to the following Honeywell SmartLine Wireless Transmitters
0-25 to 0-400 inH
2
0/0-62.2 to 0-1000 mbar Compound Characterized
0-5 to 0-100 psi/0-0.34 to 0-7 bar Compound Characterized
Honeywell SmartLine Wireless Transmitters are fully characterized at the factory and have an
applied factory calibration over the range of the transmitter (URL). Field calibration is provided
to allow up to 5% adjustment over a reduced operating range providing accuracy within .075% of
the reduce range or ±0.025 + 0.05 20 psi span psi or ±0.025 + 0.05 (1.4 bar) span bar in % span.
Compound characterized pressure transmitters allow calibration at both positive and negative
relative pressure calibration points within the characterized range of the transmitters meter body.
Table 4-2: Field calibration
SmartLine
Wireless
Model
Description
Range Limits
Span Limits
Min Lower
Range
STFW128
Flange Mounted Liquid Level
Transmitter
0 to 4.0 inH
2
O
0 to 400 inH
2
O -400 inH
2
O
STFW132
Flange Mounted Liquid Level
Transmitter
0 to 1.0 psi
0 to 100 psig
-100 psi
STFW924
Flange Mounted Liquid Level
Transmitter
0 to 10 inH
2
O
0 to 400 inH
2
O -400 inH
2
O
STFW932
Flange Mounted Liquid Level
Transmitter
0 to 5 psi
0 to 100 psig
-100 psi
STRW12D
Remote Diaphragm Seal Pressure
Transmitter
0 to 4.0 inH
2
O
0 to 400 inH
2
O -400 inH
2
O
STRW13D
Remote Diaphragm Seal Pressure
Transmitter
0 to 1.0 psi
0 to 100 psi
-100 psi
STRW93D
Remote Diaphragm Seal Pressure
Transmitter
0 to 5.0 psi
0 to 100 psig
-100 psi
Upper and Lower Calibration:
The SmartLine Wireless Transmitters provide calibration for upper and lower calibration points
as well as the option for calibration of zero. The zero calibration is independent of the upper and
lower calibration and is applied as a bias whereas upper and lower calibration adjusts the gain
and offset of the characterization. Zero calibration is available from the local interface at the
transmitter and is done with the device mounted in its operating orientation and applied “zero”
offset head pressure.
Upper and lower calibration are only available when connected to the network where access to
the upper and lower calibration points as well as access to the calibration command is available.
Upper and lower calibration requires the values for the upper and lower calibration points to be
entered and the status of the calibration operation to be verified.
Summary of Contents for SmartLine
Page 2: ......
Page 8: ......
Page 14: ...xiv SmartLine Wireless User s Manual Revision 2 ...
Page 109: ...Revision 2 SmartLine Wireless User s Manual 95 ...
Page 110: ...96 SmartLine Wireless User s Manual Revision 2 ...
Page 111: ...Revision 2 SmartLine Wireless User s Manual 97 ...
Page 119: ...Revision 2 SmartLine Wireless User s Manual 105 B5 Control Drawing ...
Page 120: ...106 SmartLine Wireless User s Manual Revision 2 ...
Page 121: ...Revision 2 SmartLine Wireless User s Manual 107 ...
Page 122: ...108 SmartLine Wireless User s Manual Revision 2 ...
Page 123: ...Revision 2 SmartLine Wireless User s Manual 109 ...
Page 128: ...114 SmartLine Wireless User s Manual Revision 2 This page has been intentionally left blank ...