Page 96
STT700 Series HART/DE Option User’s Manual
Revision 1
Reset Cold
Junction
Tracking
Values
Reset All
Method
Description
Executing this method will reset all
Cold Junction diagnostics. It resets SV
high and low values and the time
stamp, SV high and low alarm limits,
SV high and low alarm counters.
Set-up
None
NVM
Backup immediately into NVM
Reset Cold
Junction High
Value
Method
Description
Executing this method will resets Cold
Junction high value
Set-up
None
NVM
Update after every 7.5 hours.
Reset Cold
Junction Low
Value
Method
Description
Executing this method will reset Cold
Junction low value
Set-up
None
NVM
Update after every 7.5 hours.
Reset Cold
Junction High
Alarm
Counter
Method
Description
Executing this method will reset Cold
Junction high alarm counter
Set-up
None
NVM
Backup immediately into NVM
Reset Cold
Junction Low
Alarm
Counter
Method
Description
Executing this method will reset Cold
Junction low alarm counter
Set-up
None
NVM
Backup immediately into NVM