Honeywell RAE PGM-7800 Technical Note Download Page 1

Technical Note TN-151 

02/08

RAE Systems by Honeywell  877-723-2878  raesystems.com

1

ELECTROCHEMICAL SENSOR REPLACEMENT & 
MAINTENANCE

This technical note describes sensor replacement and maintenance 
procedures for electrochemical and combustible gas (LEL/TC) 
sensors used in RAE Systems instruments, including MultiRAE, 
QRAE, and VRAE. For technical specifications of sensors, see 
Technical Note TN-114, and for handling LEL sensor poisons, 
Technical Note TN-144. For CO sensor cross-sensitivity and filters, 
see Technical Note TN-121.

GENERAL

• 

Prepare a clean workspace and wash hands well before 
installing sensors. Greases and oils can cause sensors to 
perform poorly. LEL sensors are particularly sensitive to 
damage from silicone lubricants, including hand lotions.

• 

Turn off the power to the unit and remove the instrument cover.

• 

Disconnect the battery before replacing a sensor.

 
Warning:

 Failure to turn off the power during 

replacement can cause damage to some 
sensors. Failure to disconnect the battery 
can cause the non-replaceable fuse in the 
battery to blow.

Sensor Replacement Procedures

Remove the gas distribution plate, if any. Carefully pull the existing 
sensor straight out.

• 

Some electrochemical sensors are shipped with a shorting 
spring connecting two of the large pins to keep the sensor 
stabilized during storage. Remove any shorting spring before 
installing the new sensor. Replace the spring if the sensor is 
removed from the instrument for more than several minutes. 
Lack of a shorting wire will not damage the sensor, but if it 
is missing for more than about 10 minutes, the sensor may 
require an hour or more to fully equilibrate after being installed. 
Biased and LEL sensors do not have shorting springs.

• 

Insert the sensor by carefully lining up the sensor 
pins with the sockets in the circuit board. To avoid 
damaging the internal contacts, use caution and do 
not bend the pins on the sensor. Bent sensor pins can 
be carefully bent back using long-nose pliers.

• 

Do not push the sensors down too hard while 
installing, as this may cause damage to the circuit 
board components underneath the sensor.

• 

After reassembly, the unit may go into alarm if turned on 
immediately. Allow about 10 minutes for the sensor to stabilize 
in the unit before turning on the instrument. For maximum 
accuracy, non-biased toxic sensors should be allowed to 
stabilize for one hour before calibrating. Biased sensors 
require a 24-hour stabilization time with batteries installed.

• 

Re-zero and recalibrate new sensors prior to use.

 
Biased Sensors (NO)

Nitric Oxide (NO) sensors require a voltage bias and must be 
installed in position 1 of the MultiRAE Plus and positions 1 through 
3 of the VRAE, and the bias switch must be turned on. See the 
figures at the end of this Technical Note for the positions of the bias 
switches and sensors.

Warning:

 Failure to install NO sensors in the 

biased position can damage the sensors. Using 
shorting springs on bias-type sensors can 
damage the sensors.

Biased sensors require a 24-hour stabilization time with batteries 
installed. If you receive a new unit with this sensor, install the 
batteries and wait overnight before attempting to calibrate the 
sensor. Units with rechargeable batteries must be kept at minimal 
charge to main the bias on the sensor. If the sensors or batteries are 
disconnected for more than a few seconds, it may require several 
hours to again stabilize the sensor.

RECALIBRATION

The MultiRAE remembers the last calibration on a given sensor 
position. If a sensor is traded out, the new sensor needs about one 
hour to equilibrate, and it must then be recalibrated. In the case 
of a sensor with an electrical bias (NO), it must equilibrate about 

Reviews: