Security Seal Installation
Note: The certification of the weighing mechanism of the scale version of this scanner is subject to federal,
state and local Weights and Measures statutes and regulations and can only be performed by
authorized government agencies and/or their duly registered agents. Each time the scale or weighing
mechanism is calibrated, it should be properly sealed with a paper seal or a wire seal prior to being
placed into service in commerce.
It is the responsibility of the owner of the scale to confirm compliance with the relevant Weights and
Measures statutes and regulations applicable in your area by checking with the appropriate government
agency before placing a newly calibrated unit into service or removing any official seals.
Type of seal to be used will depend on the guidelines specified by the local Weights and Measures
authorities.
The security seal must only be installed if there were no errors during the scale calibration verification tests.
Pressure Sensitive Security Seal
1. Temporarily remove the platter and place it in a safe location.
2. Reinstall the calibration switch/button cover.
Figure 61. Calibration Switch/Button
Cover
3. Apply the calibration security seal over the switch cover.
Figure 62. Calibration Switch/Button
Security Seal Placement
4. Reinstall the platter.
Figure 63. Platter Installation
56
Summary of Contents for ms2420
Page 1: ...MS2420 Bar Code Scanner Avery Scale Installation and User s Guide ...
Page 10: ...6 ...
Page 13: ...Base Model Characteristics Components Figure 7 MS2420 Components 9 ...
Page 32: ...28 ...
Page 62: ...58 ...
Page 68: ...64 ...
Page 76: ...72 ...
Page 78: ...74 ...
Page 80: ...76 ...
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ...Honeywell Scanning Mobility 90 Coles Road Blackwood NJ 08012 00 05312 Rev B September 2009 ...