Calibration Mode
From the Configuration Mode, you can configure Zero, Span, Flow, and 4-20mA.
1. Go to the Monitoring Mode.
2. Press
or
buttons and hold for one second to enter the Main Menu, and one of the
Menu options is displayed.
3. Press
or
buttons to scroll to Calibration.
4. Press
to accept and enter the Calibration menu.
5. Press
or
to continue exploring more options from the Calibration Mode, and then
press
to accept it.
6. Optional Step: Press
to cancel and return one step back.
From the Calibration Mode panel, you can configure any of the following options:
Option
Operations you can calibrate from the selected option
Zero
Zero gas calibration of each channel
Span
Span gas calibration of each channel
Flow
Pump flow calibration
4-20mA
4-20mA output calibration of each channel