ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
10.1
10
S
TAINING
AND
C
OVERSLIPPING
C h a p t e r T e n
Staining and Coverslipping
GENERAL
Following is a description of
recommended guidelines
for fixation procedures, staining protocols, and
coverslipping methods.
Note:
There is wide variation among laboratories in fixation, staining, and coverslipping methods
employed for cytologic specimens. The thin layer characteristics of ThinPrep
™
processor- pre-
pared slides allow precise assessment of the effects of these differences in protocols and
allows the laboratory personnel to optimize their methods by following the general guide-
lines provided in this section. These guidelines are recommendations and should not be con-
sidered absolute requirements.
FIXATION
The ThinPrep 5000 processor with AutoLoader deposits completed slides into a staining rack
immersed in a fixative bath that contains 95% reagent alcohol or 95% ethyl alcohol. Use the following
procedure to fix ThinPrep microscope slide preparations.
•
Gyn slides:
ThinPrep microscope slides should be fixed for at least 10 minutes prior to
staining.
•
For Gyn slides intended for use with the ThinPrep
™
Imaging System:
ThinPrep
microscope slides should be fixed for at least 10 minutes prior to staining.
Note:
If the slides are being prepared for use with the ThinPrep Imaging System, please refer to the
Image Processor Operator’s Manual first.
•
Non-Gyn slides:
ThinPrep microscope slides should be fixed for at least 10 minutes prior to
staining or application of fixative spray.
Note:
Some non-gyn slides will drop into a dry bath or PreservCyt Solution, depending on the type
being run.
Change the fixative every 100 slides, or once per day, whichever comes first.
SECTION
A
SECTION
B
Summary of Contents for ThinPrep 5000
Page 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Page 5: ...This page intentionally left blank...
Page 6: ...Instructions For Use Instructions For Use...
Page 43: ...Table of Contents Table of Contents...
Page 48: ...1 Introduction 1 Introduction...
Page 73: ...2 Installation 2 Installation...
Page 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Page 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Page 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Page 134: ...6 User Interface 6 User Interface...
Page 217: ...7 Operating Instructions 7 Operating Instructions...
Page 256: ...8 Maintenance 8 Maintenance...
Page 283: ...9 Troubleshooting 9 Troubleshooting...
Page 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Page 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Page 316: ...Service Information Service Information...
Page 319: ...Ordering Information Ordering Information...
Page 326: ...Index Index...
Page 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...