Multiple Slides per Vial
Figure 7-28 Multiple Slides per Vial
Multiple slides per vial
allows you to run a non-gynecologic specimen and extract from 1 to
10 samples from the same vial. The system will bypass the fluid level too low check when processing
multiple slides per vial.
To process a sample:
1. Load a non-gynecologic sample vial into position 1 of the carousel. (Must be in position 1.)
2. Load a non-gyn filter into the filter slot and a slide into the slide slot. Load the adjacent filter
and slide slots with the number of desired samples to be made (from 2 to 10).
3. Load a filled fixative bath - with an empty slide rack - into the baths compartment.
4. Close all doors.
5. Press the
Advanced
button on the main screen.
6. Press the
Multiple slides per vial
setting button. (Note that the non-gynecologic sequence is
the only choice.) Press the green
OK
button.
The display returns to the main screen. Advanced is selected and the selected options are
detailed next to it. Figure 7-29.
7. Press the
Start
button to process the sample.
Select Multiple
slides per vial
in
order to process
up to 10 slides
out of one vial
Only Non-Gyn
sample type may
be processed with
this feature
OK
button to
return to the main
screen
SMALL BATCH MODE
SMALL BATCH MODE
Summary of Contents for ThinPrep 5000
Page 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Page 5: ...This page intentionally left blank...
Page 6: ...Instructions For Use Instructions For Use...
Page 43: ...Table of Contents Table of Contents...
Page 48: ...1 Introduction 1 Introduction...
Page 73: ...2 Installation 2 Installation...
Page 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Page 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Page 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Page 134: ...6 User Interface 6 User Interface...
Page 217: ...7 Operating Instructions 7 Operating Instructions...
Page 256: ...8 Maintenance 8 Maintenance...
Page 283: ...9 Troubleshooting 9 Troubleshooting...
Page 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Page 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Page 316: ...Service Information Service Information...
Page 319: ...Ordering Information Ordering Information...
Page 326: ...Index Index...
Page 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...