N
ON
-G
YNECOLOGIC
S
AMPLE
P
REPARATION
5.16
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
5
Procedure for the Use of DiThioThreitol (DTT) with Mucoid Non-Gyn
Samples
DTT has been shown to be a reagent that is effective in reducing the amount of mucus in respiratory
samples.
1
,
2
DTT Stock Solution
•
Prepare a stock solution by adding 2.5 g DTT
3
to 30 mL of CytoLyt
™
Solution.
•
This solution is suitable for use for 1 week when stored at room temperature (15°C–30°C).
Sample Preparation
•
This procedure is designed for mucoid non-gyn sample processing. Follow the steps for
processing mucoid specimens on the previous page.
•
After sample collection (Step 1), but prior to vortexing (Step 2), add 1 mL of the stock DTT
solution to the sample.
•
Proceed with the remaining sample processing steps as listed.
7. Allow to stand in PreservCyt Solution for 15 minutes (page 5.11)
8. Run on ThinPrep
™
5000 processor with AutoLoader using
Sequence
Non-Gyn
. Fix, stain, and evaluate.
1. Tockman, MS et al., ‘Safe Separation of Sputum Cells from Mucoid Glycoprotein’ Acta Cytologica 39, 1128
(1995).
2. Tang, C-S, Tang CMC and Kung, TM, ‘Dithiothreitol Homogenization of Prefixed Sputum for Lung Cancer
Detection’, Diagn. Cytopathol. 10, 76 (1994).
3. Available from Amresco, contact a sales representative at 800-448-4442 or
www.amresco-inc.com.
Summary of Contents for ThinPrep 5000
Page 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Page 5: ...This page intentionally left blank...
Page 6: ...Instructions For Use Instructions For Use...
Page 43: ...Table of Contents Table of Contents...
Page 48: ...1 Introduction 1 Introduction...
Page 73: ...2 Installation 2 Installation...
Page 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Page 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Page 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Page 134: ...6 User Interface 6 User Interface...
Page 217: ...7 Operating Instructions 7 Operating Instructions...
Page 256: ...8 Maintenance 8 Maintenance...
Page 283: ...9 Troubleshooting 9 Troubleshooting...
Page 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Page 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Page 316: ...Service Information Service Information...
Page 319: ...Ordering Information Ordering Information...
Page 326: ...Index Index...
Page 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...