N
ON
-G
YNECOLOGIC
S
AMPLE
P
REPARATION
5.12
ThinPrep™ 5000 Processor with AutoLoader Operator’s Manual
5
CYTOLYT SOLUTION WASH
Addition of CytoLyt
™
Solution to cell pellets is required to wash the sample. A
CytoLyt Solution
Wash
performs the following functions while preserving cellular morphology:
•
Lyse red blood cells
•
Dissolve mucus
•
Reduce protein precipitation
A
CytoLyt Solution Wash
consists of the following process:
•
Adding 30 mL of CytoLyt Solution to a cell pellet
•
Mucoid Specimens Only: Mechanical agitation
•
Concentration by centrifugation — 600g x 10 minutes
•
Pouring off the supernatant and vortexing to resuspend the cell pellet
One
CytoLyt Solution Wash
is usually adequate to clean most non-gyn samples. For particularly
bloody or mucoid specimens, additional
CytoLyt Solution Washes
may be necessary.
When a sample is collected in CytoLyt Solution at a ratio less than 30 parts CytoLyt Solution to 1 part
sample, this is considered a
Collection Step
and not a
Wash Step
. For example, if one collects 15 mL of a
sample and adds 30 mL of CytoLyt Solution to this sample, then the CytoLyt Solution: sample ratio is
only 2 to 1 and this is considered a sample collection step and still requires a
CytoLyt Solution Wash
.
For more information on CytoLyt Solution, refer to Chapter 3, PreservCyt™ & CytoLyt™ Solutions.
MUCOID
ONLY
Summary of Contents for ThinPrep 5000
Page 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Page 5: ...This page intentionally left blank...
Page 6: ...Instructions For Use Instructions For Use...
Page 43: ...Table of Contents Table of Contents...
Page 48: ...1 Introduction 1 Introduction...
Page 73: ...2 Installation 2 Installation...
Page 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Page 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Page 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Page 134: ...6 User Interface 6 User Interface...
Page 217: ...7 Operating Instructions 7 Operating Instructions...
Page 256: ...8 Maintenance 8 Maintenance...
Page 283: ...9 Troubleshooting 9 Troubleshooting...
Page 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Page 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Page 316: ...Service Information Service Information...
Page 319: ...Ordering Information Ordering Information...
Page 326: ...Index Index...
Page 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...