ADD SPECIMEN TO PRESERVCYT SOLUTION VIAL
Determine the cell pellet size and refer to the table below:
Factors to Consider
The type of pipette that you use may affect the concentration of the sample that is added to the
PreservCyt Solution vial, and therefore may affect the volume of sample. Hologic recommends using
standard, 1 mL, graduated, plastic pipettes.
If a “Sample Is Dilute” message occurs repeatedly and specimen remains in the specimen tube,
increase the number of drops of concentrated sample added to the vial.
Your technique for pouring off the supernatant may also affect the concentration of the sample. If the
supernatant is not completely poured off, then additional drops of the sample may be required. The
total volume added to the vial must not exceed 1 mL.
Size of Cell Pellet
Procedure
Pellet is clearly
visible and the
pellet volume is less
than 1 mL.
Place the centrifuge tube in a vortexor to resuspend
the cells in the residual liquid or mix the pellet by
syringing it manually with a pipette.
Transfer 2 drops of the pellet to a fresh PreservCyt
™
Solution vial.
Pellet is not visible
or is scant.
Add the contents of a fresh PreservCyt Solution vial
(20 mL) into the tube.
Vortex briefly to mix the solution and pour the entire
sample back into the PreservCyt Solution vial.
Pellet volume is
greater than 1 mL.
Add 1 mL of CytoLyt
™
Solution into the tube. Vortex
briefly to resuspend the pellet. Transfer
1 drop
of the
specimen to a fresh PreservCyt Solution vial.
Summary of Contents for ThinPrep 5000
Page 1: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...
Page 5: ...This page intentionally left blank...
Page 6: ...Instructions For Use Instructions For Use...
Page 43: ...Table of Contents Table of Contents...
Page 48: ...1 Introduction 1 Introduction...
Page 73: ...2 Installation 2 Installation...
Page 82: ...3 PreservCyt and CytoLyt Solutions 3 PreservCyt and CytoLyt Solutions...
Page 98: ...4 Gynecologic Sample Preparation 4 Gynecologic Sample Preparation...
Page 107: ...5 Non Gynecologic Sample Preparation 5 Non Gynecologic Sample Preparation...
Page 134: ...6 User Interface 6 User Interface...
Page 217: ...7 Operating Instructions 7 Operating Instructions...
Page 256: ...8 Maintenance 8 Maintenance...
Page 283: ...9 Troubleshooting 9 Troubleshooting...
Page 308: ...10 Staining and Coverslipping 10 Staining and Coverslipping...
Page 313: ...11 ThinPrep Pap Test Training Program 11 ThinPrep Pap Test Training Program...
Page 316: ...Service Information Service Information...
Page 319: ...Ordering Information Ordering Information...
Page 326: ...Index Index...
Page 345: ...Operator s Manual ThinPrep 5000 Processor with AutoLoader...