SMI-1414-EN Rev. 005 - Installation and Service Manual
Page 42 of 51
8.2 Mechanical Troubleshooting
Mechanical problems can be addressed by identifying and replacing damaged or worn components or FRUs.
Section 11 Replacement Procedures
provides guidance for components that are not replaceable by the user.
8.3 Coil Configurations
Imaging
Configuration
Connect to Left
Cable Plug
Connect to
Middle
Receptacle
Connect to
Right Cable
Plug
Configuration
to Select on
Scanner
Channels Used by
Coils
16ch Bilateral
Lateral Array
Coil Left
Medial Array
Coil
Lateral Array
Coil Right
Breast SMI
Left: 1 – 4
Medial: 5 – 12
Right: 13 – 16
8ch Left
Breast Unilateral
Lateral Array
Coil Left
Medial Array
Coil
Lateral Array
Coil Right
LtBreast SMI
Left: 1 – 4
Medial: 5 – 8
Right: Inactive
8ch Right
Breast Unilateral
Lateral Array
Coil Left
Medial Array
Coil
Lateral Array
Coil Right
RtBreast SMI
Left: Inactive
Medial: 9 – 12
Right: 13 – 16
10ch
Interventional
Bilateral
Single Loop
Coil Left
Medial Array
Coil
Single Loop
Coil Right
10Brst Int SMI
Left: 3
Medial: 5 – 12
Right: 13
5ch
Interventional
Unilateral Left
Lateral Single
Loop Coil Left
Medial Array
Coil
Lateral Single
Loop Coil Right
10BrstLt Int SMI
Left: 3
Medial: 5 – 8
Right: Inactive
5ch
Interventional
Unilateral Right
Lateral Array
Coil Left
Medial Array
Coil
Lateral Single
Loop Coil Right
10BrstRt Int SMI
Left: Inactive
Medial: 9 – 12
Right: 13
2ch Unilateral
Breast
Lateral Single
Loop Coil Left
Medial Array
Coil
Lateral Single
Loop Coil Right
2Brst Int SMI
Left: 3
Medial: Inactive
Right: 13