SMI-1414-EN Rev. 005 - Installation and Service Manual
Page 30 of 51
Figure 24: CD Contents Installation Complete
10. Exit the “Coil Data Disk Install” window and type <eject> in the original c shell. Remove the CD from
the host DVD drive.
11. To confirm the coil has been properly installed from the coil CD follow steps 2 to 4.
7.4.5 Finalization
1. Perform a SaveInfo under Guided Install to make the changes permanent.
2. Perform the Multi-Coil Quality Assurance (MCQA) test with this coil. Refer to Section 7.5.
GEHC
ONLY
3. Complete Product Locator information for serialized components. Failure do this may result in failure
of your site to receive future FMI’s. Instructions are below.
o
Go to GIB:o
After the log in screen, click on [Global Installed Base entries]
o
Under “Serial Changes to Installed Base”, enter the system ID of the system on which
this coil is being installed. Click [Submit]
o
Click on [New Entries]
- Note the PASS message.
- The lines selected show
that the Sentinelle packages
installed from the CD for a
3T system.
- The statement that “coil ID
is not in DB” is not an error
message pertaining to the
install.