Discovery QDR Series Technical Manual
4-32
Figure 4-11.
16. Click close, and then hold the Enable Lateral switch on the control panel to move
the system to the lateral position.
17. Click Start Scan to perform the lateral scan.
18. When the scan has finished, press and hold the Enable Lateral switch to return
the system to AP position.
19. Analyze the lateral scan (refer to User Guide for instructions if necessary). When
done, click Close.
20. At the Exit Analysis screen select Report. Print a report of the procedure just per-
formed, and include the report in the paperwork being returned to Hologic.
4.12 Table Top Radiographic Uniformity
This procedure verifies proper X-ray beam alignment (for Discovery A and W):
1. Restart the Discovery software in service mode (if not already).
2. Type
WBAIRQC
in the Patient Biography. Enter the serial number of the unit in
the Patient ID field.
Make sure that only the pad is on the table. Clear the table of anything else.