Discovery QDR Series Technical Manual
3-70
3.5.20 Instruments using BMD Whole Body Analysis
Instruments using only BMD Whole Body measurements can tolerate a larger S.D. If the
global S.D. is less than 3.0, you are done. If the S.D. is equal to or greater than 3.0, the
tabletop must be replaced.
Note: Archive the air scans you have acquired and either e-mail or ship them to a
Technical Support Specialist at Hologic headquarters for final evaluation.
Note: If the customer is upgrading to the Body Composition Analysis option, you must
follow the procedure and meet the requirements in “Run Reproducibility Test” on
page 3-67.
3.6 The Radiation Measurement Report
After installation, the field engineer must fill out the Radiation Measurement Report
(CSD-0042-F07) and keep this information on file.