42
Technical Data
en
11
Technical Data
Type designation
Aquilex Fluid Control System (REF: AQL-100S)
consisting of:
Irrigation Pump Unit (REF: AQL-100P)
Fluid Monitoring Unit (REF: AQL-100CBS)
Manufacturer Information
W.O.M. WORLD OF MEDICINE GmbH
Salzufer 8, 10587 Berlin
Software version
See user manual/operator instructions to determine soft-
ware version (9.3, Determine the Software Version)
Mains voltage range [V]
100-240 V~
Supply frequency range [Hz]
50/60 Hz
Fuse designation
2 x T 3,15 AH, 250 V, UL-recognized
Internal voltage supply
No
Power consumption:
Current [A]
Power consumption [VA]
Normal operation 100 V/60 Hz
0.5 A
50 VA
Normal operation 240 V/50 Hz
0.3 A
72 VA
Max. allowed load of the additional socket / multiple
sockets [A or VA]
1.6 A
Protection class (I, II, III)
I
Application part type (B, BF, CF):
BF (with Aquilex Fluid Control System inflow tube set)
1
Defibrillator protection (yes/no)
No
Protection class (IP code)
IP41 (AQL-100P), IP21 (AQL-100CBS)
Classification (I, IIa, IIb, III) in accordance with the Medi-
cal Device Directive 93/42/EEC, Appendix IX:
IIb
Conformity with the following standards:
(in the currently valid version)
IEC 60601-1 / EN 60601-1 / CAN/CSA-C22.2 NO. 60601-1 /
AAMI ANSI ES 60601-1
IEC 60601-1-2 / EN 60601-1-2 / AAMI ANSI IEC 60601-1-2
Operating conditions [°C] [°F], [%], [kPa]
10 to 40 °C/50 to 104 °F
30 to 70 % rel. humidity
70 to 106 kPa air pressure
3000 m max. altitude above sea level for device use
Possible use with explosive anesthetic gases
This device is not designed for use with flammable anes-
thetic gases (Class AP) or flammable anesthetic gases with
oxygen (Class APG).
Storage conditions [°C] [°F], [%], [kPa]
5 to 40 °C/41 to 140 °F
5 to 85 % rel. humidity
70 to 106 kPa air pressure
Transport conditions [°C] [°F], [%], [kPa]
-20 to 70 °C/-4 to 158 °F
5 to 90 % rel. humidity at 30 °C/86 °F
70 to 106 kPa air pressure
Max. sound level [dB]
< 80 dB(A) (with acoustic signals)
Max. load of container scale
25 kg/55,1 lbs, 4 containers
Max. load of bag scale
12 kg/ 26.5 lbs, 2 bags 5 l each
Maximum flow rate [l/min]
0.8 l/min ± 10 %
Adjustable values
Pressure range [mmHg]
40-150 mmHg, 5.3-20 kPa
Deficit limit [l]
0.6 - 2.5 l
Summary of Contents for Aquilex
Page 4: ......
Page 8: ......
Page 10: ......
Page 61: ...Appendix 53 en 16 Appendix 16 1 Test Log Date Tests Performed Results Comment Signature ...
Page 64: ......
Page 120: ......
Page 122: ......
Page 178: ......
Page 180: ......
Page 236: ......
Page 238: ......
Page 294: ......
Page 296: ......
Page 353: ......
Page 354: ......