Acessa
®
ProVu System User’s Guide
PL-01-0040
CO 20-131 Revision: C
9
•
To avoid damage to the electrode arrays, maintain stability of the uterus position and do not rotate the Handpiece
when electrodes are deployed in tissue.
7.
Sterilization and Safety Checks
•
Inspect the Handpiece and Pads packaging for damage.
•
Inspect the Handpiece and and Pads for any obvious damage.
•
Inspect the medical grade power cord for insulation damage or frayed wire connections.
•
Sterilize the Transducer in accordance with the recommended process in the Instructions for Use accompanying the
Acessa ProVu Transducer with sensor. Ensure that the cable interconnections are clean and dry prior to use.
•
Clean and sterilize the Handpiece Cable in accordance with the recommended process in the Instructions for Use
accompanying the Handpiece cable. Ensure that the cable interconnections are clean and dry prior to use.
•
Cl
ean and disinfect the Pad Cable in accordance with the recommended process in the Instructions for Use
accompanying the cable. Ensure that the Pad cable interconnections are clean and dry prior to use.
•
The Console, Foot Pedal, and Field Generators (TTFG) should be cleaned and disinfected per §17.3.
8.
Switches, Buttons, Connections, and Display
8.1.
Front Panel
8.2.
Buttons, Connections and Display
A.
On/Standby Button, LED Indicator
Push to turn Console ON (LED will turn green). Push again to get to the shutdown screen. The Console is in
“Ready” mode when on. If emergency RF shutoff is required during ablation, the user can turn off RF power
by momentarily pressing the front panel On/Standby button.
B.
Menu Button
Brings up the menu to access the user-adjustable settings.
C.
Menu Scroll Buttons
Allows the user to scroll through the menu items. The user-adjustable items are: full-screen ultrasound
mode, ablation volume guide on/off, coagulation power level, OR setup menu, and sound volume. The
menu will be visible when the user presses the menu button on the front console, and otherwise hidden.