background image

 

 

860D 

3 3/4 AUTO RANGE DIGIGAL DUAL DISPLAY 

  AC/DC CLAMP MULTIMETER 

OPERATION MANUAL 

This LCD Auto Range & Auto Power off Digital AC/DC clamp 

multimeter is a portable, 3 3/4-digit multimeter. It is ideally suited 
for field, laboratory, shop and home applications. 

1. SAFETY INFORMATION 

The following safety information must be observed toinsure 
maximum personal safety during the operation atthis meter. 
1) When measuring voltage ensure that instrument is not 
switched to the current range, resistance range, diode and 
continuity range, capacitance range or temperature range. 
2) Use extreme care when measuring voltage above 50V. 
especially from sources where high energy is existed. 
3) Avoid making connections to "live" circuits whenever possible. 
4) Before making resistance measurements, diode or continuity 
test, capacitance test or temperature test, ensure that the circuit 
under test isde-energized. 
5) Always ensure that the correct function and range is selected. 
6) Extreme care should be taken when using the instrument to 
conjunction with a current transformer connected to the terminals 
if an open circuit occurs. 
7) Ensure that the test leads and probes are in good condition 
with no damage to the insulation. 
8) Take care not to exceed the over-load limits as given in the 
specifications. 
9) Before opening the cover of the battery cabinet to replace 
batteries. disconnect the test leads from any external circuit, set 
the selector switch to "OFF" position. 
10) Keep the fingers after the protection ring when measuring 
through the instrument lead. 

1

1) Change the battery when the symbol “ ”  

appears  

to avoid 

incorrect data. 

 

2. Panel Layout 

 
 
 
 
 
 
 
 
 
 
 
 

 

1)

 D.HOLD/

 Back Light

 key

In any range, push the key, the 

present display value will be locked and the " H " symbol will 
appear, push it again to exit HOL and the " H"symbol disappear.

 

Press “D.HOLD” button more than 2 seconds, the back light will light, 

press it more than 2 seconds again, the back light will light off. 

2) Rotary Switch: use this switch to select functions and ranges. 
3) RANGE key

Push the key to select manual mode, push it 

again to change the range, press the key for more than 2 
seconds to go back auto range mode. But in Hz/Duty and 

Capacitance measurement, it can not select manual range mode.

 

4)

  REL  key

Push  the  key,  the  present  display  value  will  be 

stored  in  memory,  then  the  new  display  value  is  the  difference 
between input value and stored data. In Hz/Duty measurement, it 
can not work.

 

5) SELECT key:This key work on the "CAP

Ω" range only. 

Push the key to choose resistance, diode or continuity test. 
6

Hz/DUTY key

In Hz range, push the key, you can measure the 

duty,push again, go back to    Hz    measurement. In voltage 
range, push it, you can measure Hz and duty, but the 
measurement range will be smaller,and    the auto range mode 

will be changed to manual range mode. 

7) LCD Display:

 

LCD Dual Display, facilitates reads the data.

 

8)V ΩT+ Input Jack 

COM/T-    Input Jack 

9) Clamp Light easy for test the electric current in hidden place.

 

2. SPECIFICATIONS 
2.1 GENERAL SPECIFICATIONS 

Display:    3 3/4 digit LCD with a max. reading of 3999. 
Range control:    Auto range control. 
Polarity:    Automatic negative polarity indication. 
Zero adjustment: Automalic. 
Overrange indication: Only the "OL" display. 

Lowbattery: The "

" is display when the battery voltage is below 7.2V. 

Auto Power Off: 30 minutes after stopping the switch or no push 
button, the meter automatically enter to power off mode.Push 
button or run switch, auto power off disable. 
Safety Standards: The meter is up to the standards of IEC1010 
Double Insulation, Pollution Degree 2, Overvoltage Category 

.   

Clamp opening size: 65mm. 
Operating Environment: Temperature 32

104

°

F(0

40

), 

humidity

80%RH. 

Storage Environment: Temperature -4

140

°

F(-20

60

),humidity

90%RH. 

Power supply: 9V Zinc-carbon battery. 
Dimension: 275(H)×120(W)×32(D)mm. 
Weight: Approx. 562g (including batteries). 

2.2 ELECTRICAL SPECIFICATIONS 

Accuracies are ±(% of rnumder in last digit) at 23

±

5

,

70%RH. 

2.2.1 DC Voltage 

Range   

Accuracy 

Resolution 

400mV 

0.8% of rdg+5digits 

0.1mV 

4V 

1.0% of rdg+5digits 

1mV 

40V 

10mV 

400V 

100mV 

1000V 

1.5% of rdg+5 digits 

1V 

Overload protection: 1000V DC/750Vrms AC 
Impedance: 10M

Ω,

More than 100M

Ω

  on 400mV scale 

2.2.2 AC Voltage 

Range 

Accuracy 

Resolution 

Frequency 

4V 

1.5% of rdg+5 digits 

1mV 

50

400Hz 

40V 

10mV 

400V 

100mV 

750V 

2.5% of rdg+5 digits 

1V 

50

100Hz 

Average sensing, calibrated to rms of sine wave 
Overload protection: 1000V DC/750Vrms AC 
Impedance: 10M

Ω

2.2.3 DC Current 

Range 

Accuracy 

Resolution Frequency 

200A 

3.0% of rdg+10 digits 

0.1A 

3000A 

0

1000 

3.5% of rdg+20 digits 

1A 

1000

2000

  5.5% of rdg+20 digits 

2000

3000

  6.5% of rdg+40 digits 

Overload protection:3000Arms within 60 seconds 

2.2.4 AC Current

 

Range 

Accuracy 

Resolution Frequency 

200A 

3.0% of rdg+10 digits 

0.1A 

50

60Hz 

3000A 

0

1000 

3.5% of rdg+10 digits 

1A 

1000

2000

  5.5% of rdg+20 digits 

2000

3000

  6.5% of rdg+40 digits 

Average sensing, calibrated to rms of sine wave 
Overload protection: 3000Arms within 60 seconds

 

2.2.5 Resistance 

Range 

Accuracy 

Resolution 

400Ω 

1.5% of rdg+15 digits 

0.1Ω 

4k

Ω 

1.2% of rdg+15 digits 

1Ω 

40k

Ω 

10Ω 

400k

Ω 

100Ω 

4MΩ 

1k

Ω 

40MΩ 

2.5% of rdg+15 digits 

10k

Ω 

Overload protection: 250V DC/250Vrms AC 

2.2.6 Capacitance 

Range 

Accuracy 

Resolution 

40nF 

6.0% of rdg+10 digits 

10PF 

400nF 

3.5% of rdg+5 digits 

100PF 

4

μ

1nF 

40

μ

10nF 

100

μ

6.0% of rdg+10 digits 

100nF 

Overload protection: 250V DC/250Vrms AC 

2.2.7 Frequency 

Range 

Accuracy 

Resolution 

Sensitivity 

10Hz 

1.2% of rdg+10 digits 

0.01Hz 

Range of input 

voltage:1.5V~1

0V,

 

If input 

voltage over 

range,need 

adjust 

100Hz 

0.1Hz 

1000Hz 

0.8% of rdg+5 digits 

1Hz 

10kHz 

10Hz 

100kHz 

100Hz 

1000kHz 

1kHz 

10MHz  2.0% of rdg+10 digits 

10kHz 

 
Overload protection: 250V DC/250Vrms AC 

2.2.8 Temperature(NiCr-NiSi sensor) 

 