Holbach MBASS30V3 Operating Manual Download Page 1

MBASS30

V3

Mikrobiologisches Luftprobenahmesystem
M i c ro B i o l o g i c a l   A i r   S a m p l i n g   S y s t e m

Bedienungsanleitung

Operating Manual

Summary of Contents for MBASS30V3

Page 1: ...MBASS30V3 Mikrobiologisches Luftprobenahmesystem MicroBiological Air Sampling System Bedienungsanleitung Operating Manual...

Page 2: ...Das Symbol M lltonne bedeutet Batterien und Akkus d rfen nicht in den Hausm ll Die Batterien in diesem Ger t enthalten kein Blei Cadmium oder Quecksilber Note on used batteries This device contains ba...

Page 3: ...10 Configuring MBASS30V3 Setup 30 11 The Sampling Heads for MBASS30V3 33 12 The Coated Slide 43 13 Information 45 14 Service Functions 46 15 Cleaning and Maintenance Instructions 48 16 Safety Instruct...

Page 4: ...for the purposes described in this operating manual Please note the safety instructions in section 16 of this manual The following pictograms are intended to help you reading this operat ing manual Th...

Page 5: ...th the filter adapter FA 30 to load round sterile 80 mm gelatine filters Collection of antigenic and allergenic proteins in the air with the allergen sampler AS100 Comparison of the methods air sampli...

Page 6: ...volume can be preset from 10 liters to 10 000 liters Start delay from 1 second to 24 hours 4 3 inch 109 mm color display with additional glass cover Internal temperature sensor Connection socket for...

Page 7: ...remote control and adjustment of MBASS30V3 on the included data medium USB memory stick a replacement 2A slow blow fuse And depending on your ordered configuration the air sampling head LKS 30 Volume...

Page 8: ...display with touchpanel On Off button Socket for temper ature sensor Charging socket Mini USB socket D for PC connection Outlet for the sampling air USB socket H for memory stick Tripod sockets in the...

Page 9: ...common latex or nitrile gloves 4 4 Acoustic Signaler MBASS30V3 has an acoustic signaler which signals the following events 4 5 Locking Springs The three locking springs of MBASS30V3 are made of stain...

Page 10: ...ing Mechanism MBASS30V3 is equipped with a fast setting mechanism for centering the sampling media petri dishes and PS 30 slide mount By turning the setting disk clockwise the centering pins are moved...

Page 11: ...0V3 LKS100 LKS 30 PS 30 FA 30 9 4 Operating Elements Figure 5 Centering mechanism wide open disc in position 0 Figure 6 Centering mechanism narrow disc in position 12 Settings disc Recessed grips Cent...

Page 12: ...Sensor Socket MBASS30V3 features a socket for the temperature and humidity sen sor HC2A S by rotronic The temperature and humidity readings are read digitally An optional extension cable can be conne...

Page 13: ...e USB socket H host Sampling data can be exported in CSV format from the history view to this memory stick 5 4 USB Socket D MBASS30V3 can be connected to a PC running Microsoft Windows using the inclu...

Page 14: ...harger Insert the power plug into a mains socket After connecting to the mains supply the LED lights up in green Insert the plug of the quick charger into the charger socket of MBASS30V3 The state of...

Page 15: ...e charger can also be used as power supply The available current is greater than the current required for the operation of MBASS30V3 Please note the safety instructions in section 16 Figure 11 The rea...

Page 16: ...ction 7 1 Acclimate with 300l of sampling air Only for temperature differences of 10 C between MBASS30V3 and the air Prepare the sampling head LKS Section 11 1 PS Section 11 2 FA Section 11 3 Set up M...

Page 17: ...ory of MBASS30V3 Afterwards the main screen with all relevant information is shown The external temp humidity sensor is indicated by the symbol The internal temperature measurement is indicated by the...

Page 18: ...am Selecting the desired presettings is done by tapping on one of the two buttons on the left and right of the screen If the present preset programs are not as desired the parameters except for the de...

Page 19: ...1 PS 30 Section 11 2 and FA 30 Section 11 3 for this The sampling is started by pressing the Start button Note The air pressure is not shown during sampling The sampling number is increased at every s...

Page 20: ...elay time the sampling can be started by tapping the Now button During the start delay time the sampling can be aborted by tapping the Stop button 7 4 Pausing the Sampling The running sampling can be...

Page 21: ...e fan motor is electronically braked such that the additional air vol ume after the end is negligible The end of the sampling can be acknowledged by tapping the OK button The acoustic signalling lasts...

Page 22: ...voltage The air flow is blocked The preselected volume flow cannot be reached A sensor fault pressure or temperature has occurred The fan motor is faulty In these cases the sampling is terminated The...

Page 23: ...4 seconds This is to prevent aborting a sampling by accidentally pressing the button Switching off MBASS30V3 during a sampling will abort the running sampling immediately The switching off is signall...

Page 24: ...ill be given In that case recharge MBASS30V3 immediately Section 6 1 or switch it off Section 7 8 MBASS30V3 automatically turns off after 60 seconds in the low battery state 7 10 Standby Mode To save...

Page 25: ...ple the preset program no 2 will be modified such that the volume is set from 200 liters to 220 liters and the start delay from 50 seconds to 30 seconds The edit mode is started by tapping the Edit bu...

Page 26: ...s After tapping the pencil symbol the volume can be modified The current value is cleared by tapping the CLR button Afterwards the new value can be entered The new value is accepted by tapping the OK...

Page 27: ...The new value is accepted by tapping the OK button Prior to entering the new value the content of the input field is cleared with the CLR but ton To prevent input errors the entry for the volume flow...

Page 28: ...hanging the text is possible with the MBASSControl software 8 4 2 Sampling Volume The sampling volume can be set in the range from 10 liters up to 10 000 liters 8 4 3 Volume Flow The volume flow can b...

Page 29: ...n internal microSD card In the history view the data can be exported to a USB memory stick The data can also be read and di rectly be transferred into an EXCEL chart using the MBASSControl software Th...

Page 30: ...aborted alarm Sampling volume actual value Volume flow target value Sampling duration actual value Sampling start date and time Sampling volume target value Average temperature Average humidity only w...

Page 31: ...p of the start operation The file names of the sampling data end with _hist and the ones of the sampling process log with _log If there is logging data for the displayed sampling it will be exported t...

Page 32: ...ain screen Figure 13 the set tings menu is displayed Tap the blue Back button to go back to the main screen Tap the Device button to configure MBASS30V3 The first configura tion page is shown Tap the...

Page 33: ...he desired setting tap the cor responding line Save with the green button Discard with the red button 3 Time Input To change the time the input field has to be cleared first with the CLR button The ne...

Page 34: ...SS30V3 10 2 Sampling Settings In the sampling settings the number of displayed presettings programs can be set in the range from 1 to 10 To change the value the input field has to be cleared first wit...

Page 35: ...temberg DIN EN ISO 14698 1 2003 11 1 1 Functional Principle of LKS100 LKS 30 The air samplers LKS100 and LKS 30 are designed for sampling of cultivable germs Viable Microbial Sampling on culture media...

Page 36: ...ir sampler is identical with the exception that it is operated with a volume flow rate of 30 l min 11 1 2 Preparation for Sampling Remove the protective cap Open the sampler by lightly pushing the thr...

Page 37: ...ction 4 7 Remove the lid from the dish and place the top and middle parts back on MBASS30V3 Take care not to contaminate the lid of the petri dish Close LKS100 by lightly pushing the locking springs t...

Page 38: ...e petri dish with adhesive tape or another suitable sealing film Close LKS100 by lightly pushing the locking springs towards the center until the locking knows snap into place Depending on the consist...

Page 39: ...ovable slot jet permits sampling in any operating position due to the slide mounting mechanism permits sampling from hollow and tight spaces works on the slot jet impaction method is an analytical met...

Page 40: ...panol The cleaning agent can be sprayed directly through the slot jet Do not tamper with objects on the jet outlet Place the slide mount on MBASS30V3 Take the coated slide out of its shipping containe...

Page 41: ...sonable and in case of an extremely high con centration of particles often also no longer possible Overloaded sam plings often also impact on neighboring tracks which means that these also need to be...

Page 42: ...Open the sampler by lightly pushing the locking springs at the black knobs outwards Carefully remove the loaded slide and slide it into the previously opened shipping container Close the shipping cont...

Page 43: ...sche Regeln f r biologische Arbeitsstoffe Technical Regulations for Biological Materials can also be operated with the separate pump MP2 39 using the optional base unit 11 3 1 Preparation for Sampling...

Page 44: ...it according to the handling instructions of the manufacturer 11 3 3 Notes Avoid wedging FA 30 and MBASS30V3 when installing Clean FA 30 before and after each use Only use filters with a suitable flow...

Page 45: ...uction The expiry date is given in the form month year Never touch the coating of the slide The coated slides are made of glass Protect the slides against mechanical stress Only use slides with a plan...

Page 46: ...ipping container After taking the shipping con tainer out of the film tubing it can be opened by lifting the cap on one side The slide can be easily pulled out of the container after opening Don t til...

Page 47: ...rmation MBASS30V3 stores information on the operating time the drawn volume etc This information is shown under the menu item Info in the settings menu Tap the button to go back to the settings menu F...

Page 48: ...but ton in the settings menu Figure 34 The service menu opens Tap on the button Fan The slider Set can be used to manually control the fan Along with the fan speed the power is shown in percent Addit...

Page 49: ...are present a small difference pressure is measured If a limit value is exceeded the test is rated negatively 14 3 Touch Adjustment To adjust the touchpanel the touch adjustment service function is im...

Page 50: ...Do not exert pressure on the display glass 15 1 Calibration Interval The recommended calibration and maintenance interval is at every 150 operating hours of the high performance fan or at least every...

Page 51: ...SS30V3 and the charger away from children Avoid touching the display with sharp or pointed objects Do not use in areas with danger of explosion Use the charger only in dry indoor areas Do not use defe...

Page 52: ...rs Sampling medium Standard petri dish 90 mm x 15 mm diameter x height or Slide mount for particle sampler and allergen sampler Power supply LiPo battery pack 14 4 Volts 3600 mAh Buffer battery Lithiu...

Page 53: ...er for a probability of 50 of separation Round jets 324 jets each 405 m in diameter Sample air connection can be extended with a hose of 5 4 inch 31 mm inner diameter Sampling Head Air Sampler LKS100...

Page 54: ...ampling Method according to DIN ISO 16000 20 Sampling medium coated slides 76 mm x 26 mm Number of samples up to three samples per slide due to movable jet Sample air connection can be extended with a...

Page 55: ...rations of the product as well as added parts Furthermore consumable parts e g batteries fuses as well as damages caused by consumable parts e g by the leaking of batteries Also excluded is transport...

Page 56: ...ta medium MBASS30V3 is connected with the USB socket 19 1 Overview The MBASSControl software can be used to remotely control MBASS30V3 configure MBASS30V3 change the presettings programs export and tr...

Page 57: ...installa tion instructions When the software is installed for the first time on a system please make sure that the USB drivers are installed as well On subsequent installations on the same system it...

Page 58: ...the sampling But also the sampling medium must be suitable for the germination of a spore with regard to water activity substratum etc The germination of this spore can finally be hindered or even su...

Page 59: ...pores But by how much An approach to the determination of the actually collected spores is pro vided by statistics Under the assumption of a uniform distribution of the sampling air over the jet plate...

Page 60: ...particles deter mined statistically is possible only on the total spore count A projection of the CFU numbers with the individual identified species of the CFU is not possible with this statistical p...

Page 61: ...Air Sampling System MBASS30V3 Operating Manual MBASS30V3 LKS100 LKS 30 PS 30 FA 30 59 Appendix B Table for Statistical Corr of LKS 30 CFU Corr CFU Corr CFU Corr CFU Corr CFU Corr CFU Corr...

Page 62: ...System MBASS30V3 Operating Manual MBASS30V3 LKS100 LKS 30 PS 30 FA 30 60 Appendix C Table for Statistical Corr of LKS100 CFU Corr CFU Corr CFU Corr CFU Corr CFU Corr CFU Corr CFU Corr CFU Corr CFU Co...

Page 63: ...tings error Check the MBASS30V3 configura tion and correct if necessary Adjustment data error Re enter the adjustment data stated in the factory calibration certificate using the MBASSControl software...

Page 64: ...n 7 8 Air flow blocked Check that the air inlet of the sam pling head is free of obstacles Check whether the battery voltage is sufficient 20 recharge battery if necessary This error message is also d...

Page 65: ...n Spezifikationen berein The product of the declaration described above is in conformity with the requirements of the following specifications Dokument Ausgabe Documents No Date of issue Titel Title E...