Hobo RXW TEROS-12 Manual Download Page 6

RXW TEROS-12 Soil Moisture EC Temp Sensor (RXW-T12-xxx) Manual 

1-800-LOGGERS 

www.onsetcomp.com 

 

Place the mote so there is full line of sight with the next 

mote. If there is an obstruction between two sensor 

motes or between the sensor mote and the manager, then 

use a repeater mounted on the obstruction. For example, 

if there is a hill between the sensor mote and the 

manager, place a repeater at the top of the hill between 

the sensor mote and the manager. 

 

There should not be more than five motes in any direction 

at their maximum transmission range from the manager. 

Data logged by a wireless sensor must travel or “hop” 

across the wireless network from one mote to the next 

until it ultimately reaches the manager connected to the 

station. To make sure the data can successfully travel 

across the network, the mote should not be more than 

five hops away from the manager. 

 

The HOBOnet Wireless Sensor Network can support a 

maximum of 50 motes. 

 

Use a #4-40 screw to attach a ground wire to the port on 

the back of the mote if you are deploying the mote in a 

location where lightning is a concern. 

Maintenance 

If the sensor needles become contaminated with oils from 

contact with skin or another source, it is necessary to clean the 

needles to ensure accurate EC readings in salty soils with bulk 

EC greater than 10 dS/m. 
1.

 

Clean each needle using a mild detergent, such as liquid 

dish soap and a nonabrasive sponge or cloth. Avoid 

detergents that contain lotions or moisturizers. 

2.

 

Rinse the sensor and needles thoroughly with tap or 

deionized (DI) water. Do not touch the needles with an 

ungloved hand or bring them in contact with any source of 

oil or other nonconducting residue. 

The mote is designed for outdoor use, but should be inspected 

periodically. When inspecting the mote, do the following:  

 

Verify the mote is free of visible damage or cracks. 

 

Make sure the mote is clean. Wipe off any dust or grime 

with a damp cloth. 

 

Wipe off any water before opening the mote. 

 

Make sure the interior seal is intact and the latches are 

fully locked when the mote door is closed. 

Troubleshooting 

If the sensor readings are too low or slightly negative, check for 

air gaps around the sensor needles. These could be produced 

below the surface of the substrate when the needle contacts a 

large piece of material and pushes it out of the way, or if the 

sensor is not inserted perfectly linearly. 
If the sensor readings are too high, make sure the media was 

not packed excessively or insufficiently during sensor 

installation. Higher density can cause the sensor readings to be 

elevated.  

Updating Mote Firmware 

If a new firmware version is available for the mote, use 

HOBOlink to download the file to your computer. 

5.

 

In HOBOlink, go to Devices, then RX Devices, and click your 

station name. 

6.

 

On the station page, click Overview and scroll down to 

Device Information. 

7.

 

Click the Wireless tab. This icon 

 appears next to the 

mote if there is a new version of firmware available. 

8.

 

Click the firmware 

 upgrade link. Click Download and 

save the firmware .bin file to your computer. 

9.

 

Connect the mote to the computer with a USB cable (open 

the mote door and use the USB port to the right of the 

LCD). The blue LED is illuminated while connected. 

10.

 

The mote appears as a new storage device in the 

computer’s file storage manager. Copy the downloaded 

firmware file to the new storage device (the mote). The 

blue LED will blink slowly while the file is copying. 

11.

 

After the file is copied to the mote, the LED will stop 

blinking and remain a steady blue. Eject the storage device 

from the computer and disconnect the cable from the 

mote. The firmware installation process will begin 

automatically on the mote. The blue LED will blink rapidly 

while the firmware is installed. Once the firmware 

installation is complete, the LCD symbols return and the 

mote will automatically rejoin the network. 

Notes: 

 

Mac® users: A message may appear indicating the disk 

has not ejected properly when disconnecting the mote 

from the computer. The mote is operational and you can 

ignore the message. 

 

If the blue LED turns off abruptly while copying the file or 

installing the firmware, a problem has occurred. Contact 

Onset Technical Support for help. 

Battery Information 

The mote uses two 1.2 V rechargeable NiMH batteries, charged 

by the built-in solar panel. The quality and quantity of solar 

light can affect whether the battery is sufficiently charged to 

last through the night and cloudy periods. Make sure the mote 

is placed in a location that will receive several hours of sunlight 

each day. If the mote does not receive enough sunlight to 

recharge the batteries, the battery life is estimated at 3–4 

months. When batteries are regularly recharged, expected 

battery life is estimated at 3–5 years. Battery life varies based 

on the ambient temperature where the mote is deployed, the 

logging interval, the number of tripped alarms, and other 

factors. Deployments in extremely cold or hot temperatures 

can impact battery life. Estimates are not guaranteed due to 

uncertainties in initial battery conditions and operating 

environment. 
Mote operation will stop when battery voltage drops to 1.8 V. 

Mote operation will return if the battery recharges to 2.3 V. If 

the batteries are unable to be recharged, replace them with 

fresh rechargeable batteries. 

Note:

 if you install used 

rechargeable batteries that together are less than 2.3 V, the 

mote will not resume operation. 
To replace rechargeable batteries: 
1.

 

Open the mote door. 

2.

 

Remove the old batteries and install fresh ones observing 

polarity. 

Summary of Contents for RXW TEROS-12

Page 1: ...m Wireless Mote Operating Temperature Range Sensor 40 to 60 C 40 to 140 F Mote 25 to 60 C 13 to 140 F with rechargeable batteries 40 to 70 C 40 to 158 F with lithium batteries Radio Power 12 6 mW 11...

Page 2: ...he mote batteries see Deploying and Mounting Sensor Cable This is the cable that connects the mote to the sensor Eyelet Use this eyelet to attach a 3 16 inch padlock to the mote for security Latch Use...

Page 3: ...an embedded thermistor A ferrite core positioned on the sensor cable 7 6 cm 3 inches away from the sensor head is used to isolate the sensor from any interferences in the system This mitigates any pot...

Page 4: ...HOBOlink and report the mote as missing Once the mote is back online any logged data will be uploaded the next time the station connects to HOBOlink See the HOBOlink Help for details on how to change...

Page 5: ...are contaminated with oils from contact with skin or another source follow the instructions in the Maintenance section for cleaning the needles before installing the sensor Avoid any metal located bet...

Page 6: ...load the file to your computer 5 In HOBOlink go to Devices then RX Devices and click your station name 6 On the station page click Overview and scroll down to Device Information 7 Click the Wireless t...

Page 7: ...ll the new ones observing polarity 3 Push in the side tab of the solar panel cable connector and pull the connector out of the cable port 4 Place the connector in the slot on the inside of the mote do...

Page 8: ...uding interference that may cause undesired operation FCC Caution Any changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate...

Reviews: