9.
Confirm
Name of the test
or "No tests"
Confirm by pressing the
Enter
key of the DADS
Terminal, or by clicking
Cancel
button on the
main screen.
The default parameters can be changed, by changing the "Default" quality in the quality
editor application.
Quality list measurement procedure
Before tests can be done, qualities have to be defined with the Quality editor. Refer to the
manual to get more information about it.
Follow these steps to do a quality list test procedure:
Step
Action
Terminal LCD
Description
1.
Choose quality
test group.
/
Choose one of the quality test groups (by using
the group selection list in the mainframe).
2.
Choose quality
"Quality ...
Group"
Use the
Up and Down
keys of the DADS
Terminal to choose a quality out of the list. The
quality and the group name will be displayed
alternating on the LCD of the DADS Terminal.
3.
Request
measurement
start
"Quality
...Group"
Request the measurement start by pushing the
Enter
key on the DADS Terminal. Should appear
on the LCD screen.
4.
Acknowledge
"Qual. test?"
Acknowledge by pressing the
Enter
key on the
5.
Enter weight
"Weight:
xx.xx g"
Change the weight by using the
arrow keys
of the
DADS Terminal. Terminate by pressing
Enter
.
6.
Start the motor
"Start motor"
Start the motor
, by pushing the motor start
button of the Absorptometer (green lower button
on the DADS Terminal).
In case that the user pushes the
Left
key of the
DADS Terminal the test is cancelled and the
application turns back into the idle state. Start
from step 1 to redo a test.
7.
Measurement
is running
"Date"
The test name is automatically generated as date
& time and is used to file the data. The
measurement has started and the application
displays the graph torque (mNm) vs. liquid
quantity (ml/100g).
Wait for the end of test
.
Refer to Start Condition of a Measurement
Recording to get to know when the test is started
by the application.
The test can be aborted by stopping the motor.
Refer to Abort a Test for more information.