Procedure C
End-point at a fixed, but reduced torque level
If low-torque blacks are to be tested and no recording system is available, it
is recommended to reduce the setting of the torque limit switch or the
indicator set point such that the SRB F-5 gives a result of 127.0 ± 1.0 cm3 /
100 g.
This method can be simulated just by redoing the TLS measurements with
the reduced target value. This is done in the HITEC Config application, in the
. This would only be useful if data need to be correlated with
measurements from other Absorptometer without data system.
The HITEC DADS application supports the three test methods of this new method.
, to get further info on ASTM standards.
TLS Measurement
The following graph shows the TLS measurement (target value mode) of a F5 carbon black
type. The target value is 129.5 ml/100g.
Also refer to TLS Modes to get more information on the different modes the DADS can work
with.