background image

i

MN1-5821 Rev. 4

C25P Probe

Instruction Manual

Specification

MN1-5821 Rev. 4

Notes for operators and responsible maintenance personnel

★ 

Please read through this Instruction Manual as well as the separate Instruction 
Manual “Safety (MN1-5982)” and “Cleaning, Disinfection and Sterilization 
(MN1-5998)“ carefully prior to use.

★ 

Keep this Instruction Manual together with the ultrasound diagnostic instrument 
for any future reference. 

 © Hitachi, Ltd.  2014, 2017. All rights reserved.

Q1E-EP1473-4

Summary of Contents for C25P

Page 1: ...lease read through this Instruction Manual as well as the separate Instruction Manual Safety MN1 5982 and Cleaning Disinfection and Sterilization MN1 5998 carefully prior to use Keep this Instruction Manual together with the ultrasound diagnostic instrument for any future reference Hitachi Ltd 2014 2017 All rights reserved Q1E EP1473 4 ...

Page 2: ...ii MN1 5821 Rev 4 ...

Page 3: ...ds and injuries to the operator or patient Danger Indicates an imminently hazardous situation which if not avoided will result in death or serious injury to the operator or patient Warning Indicates a potentially hazardous situation which if not avoided could result in death or serious injury to the operator or patient Caution Indicates a potentially hazardous situation which if not avoided may re...

Page 4: ...parts 4 3 Preparations before use 7 3 1 Preparation for probe 7 3 1 1 Visual check 7 3 1 2 Confirmation of cleaning disinfection and sterilization 7 3 1 3 Operation check 7 3 1 4 Applicable probe cover 7 3 2 Preparation for the biopsy attachment 8 3 2 1 Inspection for the biopsy Attachment 8 3 2 1 Confrimation of the needle echo 8 4 1 How to attach a probe cover and the biopsy attachment 9 4 2 Ope...

Page 5: ...er injuries to the patient or operator 1 2 Classification of ME equipment This probe is classified as follows according to IEC60601 1 Please refer to the section 2 1 for the applied part the part treated as the applied part and the range of IPX7 Classification based on the degree of protection against electric shock Type BF applied part Classification for protection against ingress of liquids IPX7...

Page 6: ... G 10 pcs pack EZU PA7B1 C 17 G 10 pcs pack Real Time Virtual Sonography RVS Please use the option listed in Table 2 for performing RVS Regarding how to attach release the attachment please refer to the section 4 3 Table 2 Option for RVS Product Name Product No Magnetic Position Sensor Attachment RV 005 Warning Use a needle which is applicable to the biopsy attachment Use of the needle which is no...

Page 7: ...h this probe is different for each country please refer to the instrument instruction manual or contact your local Hitachi representative Field of view 70 Frequency 3 0 MHz Cable length 2 2 m Service life 3 years Applied part Probe tip including ultrasonic radiation part see the section 2 2 Part treated as Applied part Cable up to 1 m length from the probe tip IPX7 range See Figure 1 Measurement a...

Page 8: ...ach part is listed in Table 3 Connector Ultrasonic radiation part Orientation mark side Cable Probe tip Biopsy Attachment Option Needle Guide Needle Insert Biopsy Attachment Mounting Plate Probe Probe Needle Guide Needle Insert Biopsy Attachment Mounting Arm Needle Insertion Groove Figure 2 Name of each parts ...

Page 9: ...e needle insert The applicable needle size is marked on each parts Refer to 4 1 for how to attach the biopsy attachment to the probe Needle Insertion Groove A puncture needle is inserted into this part The insertion angles are 0 15 and 30 The angles are marked on the needle guide Biopsy Attachment Mounting Arm This part is used to fix the probe cover to the probe Please refer to 4 1 for how to fix...

Page 10: ... 6 MN1 5821 Rev 4 ...

Page 11: ... the ultrasound diagnostic instrument and check that the displayed scan type and frequency correspond to those of the probe Check also that there is no abnormality in the image Remark Please refer to the documentation supplied with the ultrasound diagnostic instrument for how to connect the probe and information displayed on the monitor If the probe is operated in still air brightness on the top o...

Page 12: ...e on the image Refer to the documentation supplied with the ultrasound diagnostic instrument for how to display the puncture guideline 3 Attach the biopsy attachment to the probe according to 4 1 Insert the test needle into the needle insertion groove Confirm that the needle can be smoothly inserted 4 Put the probe in sterile water so that needle echo is displayed 5 Confirm that the needle echo at...

Page 13: ...asonic radiation part Figure 4 2 Mount a probe cover on the probe tip pull the cover to smooth a wrinkle and remove air bubbles from the ultrasonic radiation part Figure 5 Be careful not to break the probe cover during this procedure 3 Place a sterile rubber band into the groove of the probe to fix the probe cover to the probe Figure 6 Ultrasonic radiation part Sterile ultrasound gel Probe cover F...

Page 14: ...ed parts of the attachment on the grooves of the probe in order to attach the biopsy attachment to the probe Note that the biopsy attachment is attached over the probe cover Figure 9 Needle Guide Needle Insert Applicable needle size Figure 7 Biopsy Attachment assembly Figure 8 Opening of Biopsy Attachment Mounting Arm Figure 9 Attachment of Biopsy Attachment Open outward Biopsy Attachment Mounting...

Page 15: ...sy attachment mounting arm to the probe while pulling the probe cover as shown in Figure 11 Note that the wrinkle at the part indicated in Figure 11 should be smoothed as possible Biopsy Attachment Slide Figure 10 Fixing Biopsy Attachment Stretch wrinkles of the probe cover Biopsy Attachment Mounting Arm Fix Figure 11 Fixing Biopsy Attachment Mounting Arm ...

Page 16: ...hed to the probe Figure 13 Visually inspect that there is no hole or tear in the probe cover Set a new probe cover if a hole or tear is found Needle Needle Confirm that the probe cover does not interfere with the needle path Figure 12 Confirmation of the needle path Attached correctly Confirm the fit of the biopsy attachment Attached incorrectly Attached incorrectly Figure 13 Confirmation of the a...

Page 17: ...r to the documentation supplied with the ultrasound diagnostic instrument Fix the probe firmly and perform a puncture by inserting a needle to the needle insertion groove Confirm that the puncture guideline which corresponds to the insertion angle is displayed Figure 14 indicates the relationship between the insertion angle and the guideline If the probe is needed to be released from the biopsy at...

Page 18: ... used Reuse of the probe cover can result in infection to the patient Caution Do not operate the probe with excessive force Use with excessive force could result in injury to the patient Scan for minimum time necessary at the lowest possible acoustic output Acoustic output may affect the patient s internal tissues For details about the acoustic output please refer to the documentation supplied wit...

Page 19: ...magnetic position sensor attachment The direction mark of the sensor is upward as shown in Figure 16 Magnetic Position Sensor Attachment Direction mark upward Magnetic Position Sensor Figure 16 How to attach Magnetic Position Sensor 2 Attach the magnetic position sensor attachment to the probe as shown in Figure 17 Figure 17 How to attach Magnetic Position Sensor Attachment Caution Attach the magn...

Page 20: ... 08 31 Manufacturer Hitachi Ltd 2 16 1 Higashi Ueno Taito ku Tokyo 110 0015 Japan Contact 81 3 6284 3668 http www hitachi com businesses healthcare index html Overseas Offices Hitachi Medical Systems GmbH Otto von Guericke Ring 3 D 65205 Wiesbaden Germany EU Importer Hitachi Medical Systems Europe Holding AG Address Sumpfstrasse 13 CH 6300 Zug Switzerland ...