39
Abort Measurement
Syntax
Query
:ABORt
Description
Aborts (forcibly terminates) the measurement or waiting for measurement completion
by
:READ?
.
Example
*TRG
:ABOR
Aborts the measurement.
:READ?
:ABOR
Aborts the waiting for measurement completion.
Query Measured Value of Centroid Wavelength (
# is R, G, or B
)
Syntax
Query
:FETCh:WAVelength:CENTroid:#?
Response
<Measured value of centroid wavelength(NR3)>,<Measurement
status(NR1)>
<Measured value of centroid wavelength> = Unit: [nm]
See:
“Measurement Value Formats” (p.37), “Measurement Status” (p.36)
Description
Reads the last measured value of the centroid wavelength.
Example
:MODE NORM
:READ?
*TRG
3.7109E-01,3.4633E-01,4.03,0
:FETC:WAV:CENT:R?
6.3427E+02,0
The red centroid wavelength is 634.27 [nm].
:FETC:WAV:CENT:G?
5.4012E+02,0
The green centroid wavelength is 540.12 [nm].
:FETC:WAV:CENT:B?
4.5208E+02,0
The blue centroid wavelength is 452.08 [nm].
Note
• # is R, G, or B.
• Be sure to check that the measurement status is an appropriate value.
• When the centroid wavelength input mode is enabled, the set centroid wavelength is the
measured value.
• When the measured value of the centroid wavelength is not within the measurable
range, the centroid wavelength is treated as unmeasured.
• If the measurement mode is not the normal measurement mode, an execution error
occurs.