28
(11) Dark Value Acquisition State
Query Dark Value Acquisition State (
# is R, G, or B
)
Syntax
Query
:DARK:STATe:#?
<Measurement range number(NRf)>
Response
<1/0(NR1)>
<Measurement range number> = 1 to 16
Description
Command
Query
Queries the dark value acquisition state.
Response is 1.: Already acquired.
Response is 0.: Not acquired. The default dark is used for the measurement.
Example
:DARK:STAT:R? 1
1
The dark value to be used for red range 1 is already acquired.
:DARK:STAT:B? 16
0
The dark value to be used for blue range 16 is not acquired.
(12) Clearing Dark Value
Clear Dark Value
Syntax
Query
:DARK:CLEar
Description
Command
Clears all of the dark data.
When the command is accepted, the measured value is cleared and the
instrument enters the non-measurement state.
Query
Note
All of the dark data are cleared and the default dark is used for the next measurement.
(13) Normal Measurement Settings
(a) Average Times
Set and Query Number of Average Times of Normal Measurement
Syntax
Command
: AVERaging
<Average times (NRf)>
Query
: AVERaging?
Response
<Average times(NR1)>
<Average times>
=
1 to 100
Description
Command
Sets the number of average times of the normal measurement.
When the command is accepted, the measured value is cleared and the
instrument enters the non-measurement state.
Query
Queries the currently set number of average times of the normal
measurement.
Example
: AVER 2
:AVER?
2